Method and system design for spectral imaging based on dual micro-lens array field-of-view segmentation.
Opt Express
; 32(8): 13293-13308, 2024 Apr 08.
Article
in En
| MEDLINE
| ID: mdl-38859303
ABSTRACT
The function of a mask in the integral field imaging spectrometer (IFIS), which segments image and samples, leads to the drawback of low spectral energy transmittance. Here, we improve field-of-view segmentation method and propose a dual micro-lens array imaging spectrometer (DMAIS). DMAIS comprises a projection lens (PL), a segmentation collimation module (SCM), and a telecentric lens (TL). And SCM, based on a dual micro-lens array, is the core component of it. By employing a lens array focusing approach instead of aperture sampling, DMAIS effectively enhances energy transmittance and reduces spectral bending. The ZEMAX simulation results indicate that compared to IFIS, DMAIS demonstrates a 109.2% increase in energy transmittance and a 32.9% reduction in spectral bending.
Full text: