Revision of metal-on-metal hip replacements with dual-mobility bearings and acetabular component retention.
Fishley, William; Nandra, Rajpal; Carluke, Ian; Partington, Paul F; Reed, Mike R; Kramer, Derek J; Wilson, Matthew J; Hubble, Matthew J W; Howell, Jonathan R; Whitehouse, Sarah L; Petheram, Timothy G; Kassam, Al-Amin M.
Affiliation
  • Fishley W; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
  • Nandra R; Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
  • Carluke I; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
  • Partington PF; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
  • Reed MR; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
  • Kramer DJ; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
  • Wilson MJ; Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
  • Hubble MJW; Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
  • Howell JR; Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
  • Whitehouse SL; Exeter Hip Group, Exeter Hip Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon University Healthcare NHS Foundation Trust, Exeter, UK.
  • Petheram TG; Queensland University of Technology, Brisbane, Australia.
  • Kassam AM; Northumbria Revision Arthroplasty Group, Northumbria Healthcare NHS Foundation Trust, Wansbeck Hospital, Ashington, UK.
Bone Jt Open ; 5(6): 514-523, 2024 Jun 24.
Article in En | MEDLINE | ID: mdl-38910515
ABSTRACT

Aims:

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods:

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).

Results:

In the DM group, two (3.8%) patients underwent further surgery one (1.9%) for dislocation and one (1.9%) for infection. In the AR group, four (9.8%) underwent further procedures two (4.9%) for loosening of the acetabular component and two (4.9%) following dislocations. There were no other dislocations in either group. In the DM group, operating time (68.4 vs 101.5 mins, p < 0.001), postoperative drop in haemoglobin (16.6 vs 27.8 g/L, p < 0.001), and length of stay (1.8 vs 2.4 days, p < 0.001) were significantly lower. There was a significant reduction in serum metal ions postoperatively in both groups (p < 0.001), although there was no difference between groups for this reduction (p = 0.674 (cobalt); p = 0.186 (chromium)).

Conclusion:

In selected patients with Birmingham MoM hips, where the acetabular component is well-fixed and in a satisfactory position with no surface damage, the metal head can be exchanged for polyethylene ADM/MDM bearings with retention of the acetabular prosthesis. This presents significant benefits, with a shorter procedure and a lower risk of complications.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Bone Jt Open Year: 2024 Document type: Article Affiliation country: Reino Unido

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Bone Jt Open Year: 2024 Document type: Article Affiliation country: Reino Unido
...