Your browser doesn't support javascript.
loading
Viscoelastic, Optical, and Surgical Properties of Vitreous Body Replacement Hydrogels After Aging Compared to Porcine Vitreous Bodies And Silicone Oils.
Hammer, Maximilian; Muuss, Marcel; Herbster, Lorenz; Herth, Jonathan; Scheuerle, Alexander; Khoramnia, Ramin; Labuz, Grzegorz; Uhl, Philipp; Auffarth, Gerd Uwe.
Affiliation
  • Hammer M; David J Apple Laboratory for Vision Research, Heidelberg, Germany.
  • Muuss M; University Eye Clinic Heidelberg, Heidelberg, Germany.
  • Herbster L; David J Apple Laboratory for Vision Research, Heidelberg, Germany.
  • Herth J; University Eye Clinic Heidelberg, Heidelberg, Germany.
  • Scheuerle A; Institute for Pharmacy and Molecular Biotechnology, Heidelberg, Germany.
  • Khoramnia R; David J Apple Laboratory for Vision Research, Heidelberg, Germany.
  • Labuz G; University Eye Clinic Heidelberg, Heidelberg, Germany.
  • Uhl P; University Eye Clinic Heidelberg, Heidelberg, Germany.
  • Auffarth GU; David J Apple Laboratory for Vision Research, Heidelberg, Germany.
Transl Vis Sci Technol ; 13(7): 5, 2024 Jul 01.
Article in En | MEDLINE | ID: mdl-38967936
ABSTRACT

Purpose:

First- (monomers), second- (pre-gelated), and third- (in situ gelating after injection) generation hydrogels were previously introduced to replace the vitreous body after vitrectomy surgery. In this study, we evaluated the surgical, optical, and viscoelastic properties of vitreous body replacement hydrogels before and after an accelerated aging protocol previously applied to intraocular implants.

Methods:

Measurements of injection force, removal speed using a clinically established vitrectomy setup, as well as evaluation of forward light scattering and viscoelastic properties before and after an accelerated aging protocol were conducted. Results were compared to porcine and human vitreous bodies, as well as currently clinically applied lighter- and heavier-than-water silicone oils.

Results:

Removal speed of all tested hydrogels is substantially lower than the removal speed of porcine vitreous body (0.2 g/min vs. 2.7 g/min for the best performing hydrogel and porcine vitreous body, respectively). Forward light scattering in second-generation vitreous body replacement hydrogels was higher after the aging process than the straylight of the average 70-year-old vitreous body (9.4 vs. 5.5 deg2/sr, respectively). The viscoelastic properties of all hydrogels did not change in a clinically meaningful manner; however, trends toward greater stiffness and greater elasticity after aging were apparent.

Conclusions:

This study demonstrates surgical weaknesses of the hydrogels that need to be addressed before clinical use, especially low removal speed. Pre-linked hydrogels (second-generation) showed inferior performance regarding surgical properties compared to in situ gelating hydrogels (third-generation). Translational Relevance This study highlights possible pitfalls regarding surgical and optical properties when applying vitreous replacement hydrogels clinically.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Vitrectomy / Vitreous Body / Silicone Oils / Hydrogels Limits: Aged / Animals / Humans Language: En Journal: Transl Vis Sci Technol Year: 2024 Document type: Article Affiliation country: Alemania

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Vitrectomy / Vitreous Body / Silicone Oils / Hydrogels Limits: Aged / Animals / Humans Language: En Journal: Transl Vis Sci Technol Year: 2024 Document type: Article Affiliation country: Alemania