A high-sensitivity electrochemiluminescence-based detection system for automated PCR product quantitation.
DiCesare, J; Grossman, B; Katz, E; Picozza, E; Ragusa, R; Woudenberg, T.
Affiliation
  • DiCesare J; Perkin-Elmer Corporation, Wilton, CT 06897.
Biotechniques ; 15(1): 152-7, 1993 Jul.
Article in En | MEDLINE | ID: mdl-8363832
ABSTRACT
A high-sensitivity nonisotopic system has been developed for post-PCR product detection. The probe-based detection system exploits a chemiluminescent reaction that takes place on the electrode surface in an electrochemical cell. The detection system incorporates a biotin-streptavidin capture reaction onto a solid support that permits fast post-PCR product detection at the attomole level. The system precision is within 5% relative standard deviation over a linear dynamic range of greater than three orders of magnitude. In this paper, the principles and features of the electrochemiluminescent-based detection system, together with its application to PCR product quantitation, are described in detail.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Autoanalysis / Polymerase Chain Reaction / Luminescent Measurements Type of study: Diagnostic_studies Limits: Humans Language: En Journal: Biotechniques Year: 1993 Document type: Article