ABSTRACT

BACKGROUND:

We developed an analytical method to measure alpha-L-iduronidase (IDUA) activity in dried blood spots. This was achieved by using liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) with electrospray ionization in the positive ion mode.

METHODS:

Chromatographic separation was completed using mobile phase involving water-formic acid and acetonitrile-formic acid over 2.8 min of run time on a column with a Kinetex XB-C18 (Phenomenex, USA). The detection of column effluent was performed using a Xevo TQ-S triple quadrupole mass spectrometer (Waters, USA) in the multiple-reaction monitoring mode. This method was verified with blank and control samples at four activity levels base, low, medium, and high. Control materials were provided from Centers for Disease Control and Prevention (CDC).

RESULTS:

Intra- and inter-day precisions were between 2.6% and 16.5% and between 7.9% and 17.0%, respectively. A correlative regression study on the IDUA activity in CDC-control samples performed to assess the validity of the developed method showed a highly significant linear association (r2=0.9976) between the calculated and CDC-reported values and an obvious difference in activity among the four levels. This reliable analytical method was applied to mucopolysaccharidosis I (Hurler) screening of patients under treatment (n=4) and in normal controls (n=129). IDUA activity ranged from 8.98 to 77.12 micromol/hr/L) in normal controls, and patients undergoing medical treatment showed low IDUA activity.

CONCLUSIONS:

This method had advantages of simplicity, rapid sample preparation, and liquid chromatographic separation, which efficiently inhibited ionization suppression induced by matrix effects in mass spectrometric detection.
Subject(s)
Key words
Full text: 1 Database: WPRIM Main subject: Substrate Specificity / Regression Analysis / Chromatography, High Pressure Liquid / Mucopolysaccharidosis I / Tandem Mass Spectrometry / Dried Blood Spot Testing / Iduronidase Type of study: Diagnostic_studies / Screening_studies Limits: Humans Language: En Journal: Annals of Laboratory Medicine Year: 2015 Document type: Article
Full text: 1 Database: WPRIM Main subject: Substrate Specificity / Regression Analysis / Chromatography, High Pressure Liquid / Mucopolysaccharidosis I / Tandem Mass Spectrometry / Dried Blood Spot Testing / Iduronidase Type of study: Diagnostic_studies / Screening_studies Limits: Humans Language: En Journal: Annals of Laboratory Medicine Year: 2015 Document type: Article
...