Dosimetric verification for radiotherapy quality audit under reference and non-reference conditions in Jiangsu province / 中华放射医学与防护杂志
Article in Zh | WPRIM | ID: wpr-453536
Responsible library: WPRO
ABSTRACT
Objective To verify the methodology for auditing dosimetric parameters in reference and non-reference conditions with thermoluminescent dosimeters (TLDs).Methods Under reference and non-reference conditions,the established TLD methods were used to observe the absorbed dose variations with depth,SSD,field size and 45 wedges for 10 photon beams at 5 hospitals.Dosimetric parameters,including doses at Dmax points in axis,on 5 electron beams of 9 MeV were measured.The measurement results were compared between the TLDs and plane parallel ionization chambers.Results For 6 MV photon beams,the relative deviation of between finger ionization chamber method and TLD chips was in the range of-1.7% to 5.4% under on-axis non-reference conditions,and-6.3% to-0.6% under off-axis non-reference conditions,respectively,all within the range of ≤ ± 7% as required by the IAEA.The relative deviation between plane parallel chamber and TLD method was-2.3% to 3.7%,within ± 5% as required by the IAEA.Conclusions It is convenient and feasible to use TLD method for quality audits of dosimetric parameters in radiotherapy.
Key words
Full text: 1 Database: WPRIM Language: Zh Journal: Chinese Journal of Radiological Medicine and Protection Year: 2014 Document type: Article