Evaluation of the Self-Testing Blood Glucose Monitoring System GlucoDr.S According to ISO 15197:2013 Guidelines
Article in Ko | WPRIM | ID: wpr-715913
Responsible library: WPRO
ABSTRACT
BACKGROUND: The performance of the self-monitoring of blood glucose in patients with diabetes should be properly evaluated to ensure strict glycemic control. This study evaluated the self-testing Blood Glucose Monitoring System GlucoDr.S™ (All Medicus Co., Ltd., Korea). METHODS: This study recruited 120 patients. Use of the glucometer was evaluated according to ISO 15197:2013 guidelines. The YSI 2300 STAT PLUS Glucose Analyzer (YSI Life Sciences, USA) was used as the reference device. RESULTS: The standard deviation and coefficients of variation ranges for measurement repeatability and intermediate measurement precision conducted with 10 meters and 3 reagent lots on the same day were 2.7–3.2 mg/dL (0.99. The influence effect of hematocrit and the 24 interference agents was not significant, except for xylose. A system accuracy test was conducted with 100 subjects taking duplicate measurements from each of the 3 reagent lots. When glucose levels were 95% of the samples were within ±15 mg/dL and within ±15% of the average measured values of the reference measurement, respectively. In Consensus Error grid analysis, all results were distributed in zone A and B. The results of the user performance evaluation using 115 lay persons were also included in the acceptance range. CONCLUSION: The GlucoDr.S™ showed acceptable performance according to the ISO 15197:2013 guidelines and could be a clinically useful self-testing glucometer.
Subject(s)
Key words
Full text: 1 Database: WPRIM Main subject: Xylose / Blood Glucose / Biological Science Disciplines / Consensus / Glucose / Hematocrit Type of study: Guideline Limits: Humans Language: Ko Journal: Laboratory Medicine Online Year: 2018 Document type: Article
Full text: 1 Database: WPRIM Main subject: Xylose / Blood Glucose / Biological Science Disciplines / Consensus / Glucose / Hematocrit Type of study: Guideline Limits: Humans Language: Ko Journal: Laboratory Medicine Online Year: 2018 Document type: Article