Your browser doesn't support javascript.
loading
Diseño y desarrollo de un modelo experimental para estudios biomecánicos de fracturas tipo B de Tile del anillo pélvico / Design and development of an experimental model for biomechanical studies of Tile type B1 fracture of the pervic ring
Serrano Escalante, F. J; Giráldez Sánchez, M. A; Cano Luis, P; García Rodríguez, J; Navarro Robles, A; Carranza Bencano, A.
Affiliation
  • Serrano Escalante, F. J; Hospital Universitario Virgen del Rocio. Sevilla. España
  • Giráldez Sánchez, M. A; Hospital Universitario Virgen del Rocio. Sevilla. España
  • Cano Luis, P; Hospital Universitario Virgen del Rocio. Sevilla. España
  • García Rodríguez, J; Escuela Superior Ingenieros. Sevilla. España
  • Navarro Robles, A; Escuela Superior Ingenieros. Sevilla. España
  • Carranza Bencano, A; Hospital Universitario Virgen del Rocio. Sevilla. España
Trauma (Majadahonda) ; 22(3): 174-183, jul.-sept. 2011.
Article in Es | IBECS | ID: ibc-91020
Responsible library: ES1.1
Localization: BNCS
RESUMEN

Objetivo:

Diseñar un modelo estandarizado y reproducible para la realización de ensayos biomecánicos de fracturas tipo B1 de Tile del anillo pélvico.

Metodología:

Establecida en dos fases. Fase 1, determinación de requerimientos fundamentales del sistema. Fase 2, diseño, testado y elección de las piezas del modelo experimental. Se empleó una máquina universal de ensayos (MTS- 810), programas de diseño gráfico (Solid Edge Academic), fantomas, rail de carga deslizante y materiales industriales de Fabricación del Sistema de Fijación.

Resultados:

Se obtuvo un modelo constituido por dos sistemas de anclaje, superior e inferior, que permiten articular la pelvis a la máquina de ensayo tanto proximalmente, a través del sacro y L5, como a nivel distal con ambos fémures. El sistema permitió fijación adecuada de especímenes a máquina de ensayo, mantener la estática postural de bipedestación a la carga y reproductibilidad de lesión B1 de Tile.

Conclusión:

Es posible diseñar y fabricar un sistema estandarizado de fijación de pelvis humanas a una máquina de análisis de materiales para el estudio biomecánico de las fracturas tipo B1 de Tile del anillo pélvico (AU)
ABSTRACT
Objetive The aim of the study is to design a standardized and reproducible procedure for biomechanical testing of Tile B1 fractures of the pelvic ring.

Methodology:

The methodology for this study was divided in two phases. In the first phase, the requirements to be covered by the system were determined, and in the second one experimental model pieces were designed and chosen. An Universal Testing Machine (MTS-810), Graphic Design Plans Program Software (Solid Edge Academic), Fantomas, Bearing, Sliding Rail and Manufacturing Material Fixing System were used.

Results:

A specific loading biomechanic system has been created, consisting on an upper and a lower anchor, which allows pelvis ring fix to the test machine both proximally through the sacrum and the last lumbar vertebra, and in terms of both femurs to the database of the testing machine. The procedure make a correct bone fixation, place the pelvis in a stand position, getting to align the upper and lower anterior iliac spines in the same plane and reproduce Tile B1 fracture.

Conclusion:

It is possible to design and manufacture a simple standard system for determining human pelvis to a testing machine for biomechanical studies of Tile type B1 pelvic ring fractures (AU)
Subject(s)
Search on Google
Collection: 06-national / ES Database: IBECS Main subject: Pelvis / Prostheses and Implants / Biomechanical Phenomena / Implants, Experimental / Suture Anchors Type of study: Prognostic_studies Language: Es Journal: Trauma (Majadahonda) Year: 2011 Document type: Article
Search on Google
Collection: 06-national / ES Database: IBECS Main subject: Pelvis / Prostheses and Implants / Biomechanical Phenomena / Implants, Experimental / Suture Anchors Type of study: Prognostic_studies Language: Es Journal: Trauma (Majadahonda) Year: 2011 Document type: Article