Your browser doesn't support javascript.
loading
Braquiterapia endobrônquica de alta taxa de dose: estudo dosimétrico / High dose-rate endobronchial brachytherapy: a dosimetric study
Lopes, Regina Maria Godoy; Rodrigues, Laura Natal; Carvalho, Heloisa de Andrade.
Affiliation
  • Lopes, Regina Maria Godoy; s.af
  • Rodrigues, Laura Natal; s.af
  • Carvalho, Heloisa de Andrade; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Serviço de Radioterapia. São Paulo. BR
Radiol. bras ; 43(1): 1-6, jan.-fev. 2010. ilus, tab
Article in Portuguese | LILACS | ID: lil-542681
Responsible library: BR734.1
RESUMO

OBJETIVO:

Avaliar a distribuição de dose em diferentes situações de braquiterapia endobrônquica de alta taxa de dose, com foco principalmente nos volumes de altas doses, e tentar definir situações de melhor ou pior distribuição de dose que possam servir de guia na prática clínica.

MATERIAIS E MÉTODOS:

Estudo teórico, simulando braquiterapia endobrônquica de alta taxa de dose utilizando dois cateteres, com variação da extensão de carregamento, angulação entre os cateteres, profundidade de cálculo e o intervalo entre as paradas da fonte. Com prescrição de 7,5 Gy, foram calculados os volumes englobados pelas isodoses correspondentes a 100 por cento, 150 por cento e 200 por cento da dose prescrita (V100, V150 e V200, respectivamente) e as razões V150/V100 e V200/V100.

RESULTADOS:

Os volumes aumentaram com o aumento da extensão de carregamento dos cateteres, profundidade de cálculo e angulação, com tendência a um aumento proporcionalmente menor para angulações maiores. As relações V150/V100 e V200/V100 foram, em geral, homogêneas, ao redor de 0,50 e 0,30, respectivamente.

CONCLUSÃO:

A distribuição de dose na situação considerada padrão é em geral adequada. Nenhum parâmetro específico que pudesse ser relacionado à maior toxicidade foi identificado. Recomendamos uma avaliação rápida da qualidade do implante por meio da análise das relações V150/V100 e V200/V100.
ABSTRACT

OBJECTIVE:

To evaluate the dose distribution in different situations of high dose-rate endobronchial brachytherapy, focusing especially on high-dose volumes, and try to identify better or worse situations in terms of dose distribution to aid as guidance in the clinical practice. MATERIALS AND

METHODS:

Theoretical study simulating high dose-rate endobronchial brachytherapy utilizing two catheters, varying the loading extent, angle between the catheters, prescription depth, and source step. With a prescription dose of 7.5 Gy, the volumes involved by the 100 percent, 150 percent and 200 percent isodoses (V100, V150 and V200, respectively) and V150/V100 and V200/V100 ratios were calculated.

RESULTS:

There was a volume enhancement with larger loaded lengths, increase in prescription depth and angles, with a tendency towards a proportionally smaller increase with larger angulations. In general, the V150/V100 and V200/V100 ratios were homogeneous, respectively around 0.50 and 0.30.

CONCLUSION:

Overall, the dose distribution in the standard situation was appropriate. No specific parameter that could be related to a higher toxicity was identified. The authors recommend a swift evaluation of the treatment quality through the analysis of the V150/V100 and V200/V100 ratios.
Subject(s)

Full text: Available Collection: International databases Database: LILACS Main subject: Radiation Dosage / Brachytherapy / Dosimetry Type of study: Practice guideline / Incidence study Limits: Humans Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2010 Document type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade de São Paulo/BR
Full text: Available Collection: International databases Database: LILACS Main subject: Radiation Dosage / Brachytherapy / Dosimetry Type of study: Practice guideline / Incidence study Limits: Humans Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2010 Document type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade de São Paulo/BR
...