Estudo da dose nos exames de tomografia computadorizada abdominal em um equipamento de 6 cortes / Dosimetry in abdominal imaging by 6-slice computed tomography
Radiol. bras
; 45(6): 326-333, out.-dez. 2012. ilus, tab
Article
in Pt
| LILACS
| ID: lil-660793
Responsible library:
BR734.1
RESUMO
OBJETIVO:
Determinar a dose efetiva recebida nos exames de tomografia computadorizada abdominal e estudar a influência das características dos pacientes na dose recebida.MATERIAIS E MÉTODOS:
Foram realizadas medições dos valores de dose com uma câmara de ionização em objetos simuladores, de forma a verificar se os valores obtidos estavam de acordo com os valores apresentados pelo equipamento de tomografia computadorizada e se estes não ultrapassavam os níveis de referência de dose recomendados. Posteriormente, foram medidos os valores de dose recebida pelos pacientes, com autonomia física, nos exames de tomografia computadorizada abdominal (n = 100) e a relação existente com as suas características antropométricas. Por último, foi simulada a dose nos órgãos por meio do método de Monte Carlo utilizando o software de simulação CT-Expo V 1.5, e estudado o efeito do controle automático de exposição nestes exames.RESULTADOS:
As principiais características com influência direta na dose são a massa corporal, o perímetro abdominal e o índice de massa corporal do paciente, cuja relação é linear e positiva.CONCLUSÃO:
A dose de radiação recebida nos exames abdominais depende de algumas características dos pacientes, sendo importante ajustar os parâmetros de aquisição às suas dimensões.ABSTRACT
OBJECTIVE:
To determine the effective dose in abdominal computed tomography imaging and to study the influence of patients' characteristics on the received dose. MATERIALS AND METHODOS Dose values measurements were performed with an ionization chamber on phantoms to check the agreement between dose values and those presented by the computed tomography apparatus, besides their compliance with the recommended reference dose levels. Later, values of dose received by physically able patients submitted to abdominal computed tomography (n = 100) were measured and correlated with their anthropometric characteristics. Finally, the dose to organs was simulated with the Monte Carlo method using the CT-Expo V 1.5 software, and the effect of automatic exposure control on such examinations.RESULTS:
The main characteristics directly influencing the dose include the patients' body mass, abdominal perimeter and body mass index, whose correlation is linear and positive.CONCLUSION:
The radiation dose received from abdominal CT scans depends on some patient's characteristics, and it is important to adjust the acquisition parameters to their dimensions.Key words
Full text:
1
Collection:
01-internacional
Database:
LILACS
Main subject:
Anthropometry
/
Dosimetry
Type of study:
Health_economic_evaluation
Limits:
Adolescent
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
Language:
Pt
Journal:
Radiol. bras
Journal subject:
DIAGNOSTICO POR IMAGEM
/
RADIOLOGIA
Year:
2012
Document type:
Article
Affiliation country:
Portugal
Country of publication:
Brazil