Your browser doesn't support javascript.
loading
Estimativa dos valores de MSAD em procedimentos de tomografia computadorizada utilizando filmes radiocrômicos / Estimation of MSAD values in computed tomography scans using radiochromic films
Oliveira, Bruno Beraldo; Mourão, Arnaldo Prata; Silva, Teógenes Augusto da.
Affiliation
  • Oliveira, Bruno Beraldo; s.af
  • Mourão, Arnaldo Prata; Centro Federal de Educação Tecnológica de Minas Gerais. Belo Horizonte. BR
  • Silva, Teógenes Augusto da; s.af
Radiol. bras ; 46(2): 96-100, mar.-abr. 2013. ilus, graf, tab
Article in Portuguese | LILACS | ID: lil-673352
Responsible library: BR734.1
RESUMO

OBJETIVO:

Verificar a viabilidade de filmes radiocrômicos como um dosímetro alternativo para estimativa da dose média em cortes múltiplos a partir dos perfis de kerma.

MATERIAIS E MÉTODOS:

Os filmes foram distribuídos em cilindros posicionados no centro e nas regiões periféricas de um objeto simulador padrão de abdome utilizado para dosimetria em tomografia computadorizada.

RESULTADOS:

Os valores de dose média em cortes múltiplos calculados foram 13,6 ± 0,7, 13,5 ± 0,7 e 18,7 ± 1,0 mGy para os valores de passo (pitch) de 0,75, 1,00 e 1,50, respectivamente.

CONCLUSÃO:

Apesar de os resultados mostrarem valores menores que o nível de referência de radiodiagnóstico de 25 mGy estabelecido pela legislação brasileira para exames de abdome, eles sugerem que há espaço para otimização dos procedimentos e uma revisão do valor para o nível de referência de radiodiagnóstico brasileiro.
ABSTRACT

OBJECTIVE:

To evaluate the feasibility of using radiochromic films as an alternative dosimeter to estimate the multiple scan average dose on the basis on kerma profiles.MATERIALS AND

METHODS:

The radiochromic films were distributed in cylinders positioned in the center and in four peripheral bores of a standard abdominal phantom utilized for computed tomography dosimetry.

RESULTS:

Values for multiple scan average dose values corresponded to 13.6 ± 0.7, 13.5 ± 0.7 and 18.7 ± 1.0 mGy for pitch of 0.75, 1.00 and 1.50, respectively.

CONCLUSION:

In spite of results showing lower values than the reference level for radiodiagnosis (25 mGy) established by the Brazilian regulations for abdominal studies, it is suggested that there is room to optimize procedures and review the reference level for radiodiagnosis in Brazil.
Subject(s)


Full text: Available Collection: International databases Database: LILACS Main subject: Dosimetry / Kerma Type of study: Practice guideline Limits: Female / Humans / Male Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2013 Document type: Article Affiliation country: Brazil Institution/Affiliation country: Centro Federal de Educação Tecnológica de Minas Gerais/BR

Full text: Available Collection: International databases Database: LILACS Main subject: Dosimetry / Kerma Type of study: Practice guideline Limits: Female / Humans / Male Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2013 Document type: Article Affiliation country: Brazil Institution/Affiliation country: Centro Federal de Educação Tecnológica de Minas Gerais/BR
...