Desarrollo de una formulación estable de ureasa líquida para la determinación de urea en suero / Development of a stable formulation of liquid urease for determination of serum urea
García Borges, Lisandra; Rodríguez Fernández, Belkis; Aja Maza, Gissel; Pérez Carrasco, Juan Alberto; Russeaux Guía, Niurka; Valdés Diez, Lilliam; Licea Tornés, María Victoria.
Affiliation
  • García Borges, Lisandra; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Rodríguez Fernández, Belkis; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Aja Maza, Gissel; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Pérez Carrasco, Juan Alberto; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Russeaux Guía, Niurka; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Valdés Diez, Lilliam; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
  • Licea Tornés, María Victoria; Empresa de Producción de Biológicos Carlos J. Finlay. La Habana. CU
Rev. cuba. farm ; 48(3)jul.-set. 2014. Ilus, tab
Article in Spanish | LILACS, CUMED | ID: lil-740914
Responsible library: CU1.1
RESUMEN

INTRODUCCIÓN:

la determinación de urea en suero reviste gran importancia para el diagnóstico clínico de diversas afecciones de origen renal. Los métodos más aplicados para este fin implican el uso de la enzima ureasa por su alta especificidad en la hidrólisis de este analito, entre los cuales el método de Berthelot, emplea la ureasa en el primer paso de la reacción, seguido de una reacción colorimétrica del amoníaco liberado. La presentación de la ureasa en forma líquida es uno de los elementos que hacen más sencillo este proceso, de fácil y rápida ejecución con resultados confiables.

OBJETIVO:

desarrollar una solución de ureasa estable, como componente del juego de diagnosticador de urea en suero.

MÉTODOS:

se diseñó y optimizó la formulación teniendo en cuenta la concentración de la enzima, las soluciones buffer, los preservos y los compuestos orgánicos polihidroxilados para lograr un producto en forma líquida estable. Se evaluó la especificidad, precisión, linealidad, exactitud (comparación de métodos) y la sensibilidad del método; así como la estabilidad en vida de estante de 2 a 8 °C, por un período de 18 meses.

RESULTADOS:

se logró un reactivo líquido estable (> 12 meses de 2 a 8 °C), donde el método analítico resultó ser específico y lineal hasta 30 mmol/L de urea (r2= 0,999), sensible, preciso coeficiente de variación, CV< 3 por ciento) y exacto (r 0,999), satisfactorio para el uso al que se destina el producto, con calidad analítica comparable con los existentes en el mercado.

CONCLUSIONES:

el desarrollo de la formulación de ureasa líquida estable, permitirá introducir este componente esencial, en el juego de reactivos de urea que oferta la Industria Nacional al Sistema de Salud de acuerdo con los requisitos vigentes (Reg. 8-2001) para su comercialización(AU)
ABSTRACT

INTRODUCTION:

determination of serum urea has great importance for the clinical diagnosis of several renal illnesses. The most implemented methods involve the use of the urease enzyme because of its high specificity for this analyte hydrolysis; the Berthelot method uses urease in the first reaction step followed by colorimetric reaction of the released ammonium. The presentation of urease in liquid form is one of the elements that make this process simpler, of easy and fast application with reliable results.

OBJECTIVE:

to develop a stable urease solution as part of the diagnostic set for serum urea.

Methods:

THE ADEQUATE FORMULATION WAS DESIGNED AND OPTIMIZED BY TAKING INTO account the enzyme concentration, the buffer solutions, the preserves and the polyhydroxyled organic compounds to attain a final stable liquid product. The parameters called specificity, precision, linearity, accuracy (comparison of methods) and sensitivity of the method were all evaluated in addition to the shelf life at 2 a 8 °C for 18 months.

RESULTS:

astable liquid reagent (over 12 months at 2 a 8 °C ) was reached; the analytical method proved to be specific and linear up to 30 mmol/L of urea (r2= 0,999), sensitive, precise, with variation coefficient lower than 3 percent and accurate ((r 0,999), satisfactory for the intended use of the product and the analytical quality comparable to that of the already existing products.

CONCLUSIONS:

the development of the stable liquid urease formulation will allow introducing this essential component in the reagent set for urea offered by the national industry to the healthcare system according to the requirements for commercialization (Reg 8-2001)(AU)
Subject(s)

Full text: Available Collection: International databases Database: CUMED / LILACS Main subject: Clinical Diagnosis / Clinical Laboratory Techniques Type of study: Diagnostic study Limits: Humans Language: Spanish Journal: Rev. cuba. farm Journal subject: Pharmacy Year: 2014 Document type: Article Affiliation country: Cuba Institution/Affiliation country: Empresa de Producción de Biológicos Carlos J. Finlay/CU
Full text: Available Collection: International databases Database: CUMED / LILACS Main subject: Clinical Diagnosis / Clinical Laboratory Techniques Type of study: Diagnostic study Limits: Humans Language: Spanish Journal: Rev. cuba. farm Journal subject: Pharmacy Year: 2014 Document type: Article Affiliation country: Cuba Institution/Affiliation country: Empresa de Producción de Biológicos Carlos J. Finlay/CU
...