Performance of the platelet function analyser PFA-100 in testing abnormalities of primary haemostasis.
Blood Coagul Fibrinolysis
; 10(1): 25-31, 1999 Jan.
Article
in En
| MEDLINE
| ID: mdl-10070832
ABSTRACT
The PFA-100 device is a new instrument for the in-vitro testing of platelet function. Primary haemostasis is stimulated by recording the closure time taken for platelets to seal a 150 microm aperture in the centre of a membrane coated with collagen and either epinephrine or ADP. Patients with type 3 von Willebrand's disease (n = 4) all had infinitely prolonged closure times (> 200 s) with both types of cartridge. A patient with afibrinogenemia exhibited only slightly prolonged closure times of 111 and 166 s for the ADP and epinephrine membranes, respectively. Patients with Glanzmann's thrombasthenia (n = 6) and Bernard Soulier syndrome (n = 2) had grossly prolonged closure times (> 200 s) with both types of cartridges. These results confirmed that the PFA-100 system was highly dependent on normal von Willebrand factor, glycoprotein Ib and glycoprotein IIb/IIIa levels but not on plasma fibrinogen. Patients with storage pool disease (n = 6) and Hermansky Pudlak syndrome (n = 7) had prolonged closure times with the epinephrine cartridge. There was no evidence of enhanced platelet function in patients with antiphospholipid syndrome, in sickle-cell disease or thalassemia. However, ingestion of aspirin resulted in a near consistent and significant prolongation of the closure time for the epinephrine cartridge but not for the ADP cartridge in both normal subjects and patients. The test offers a reliable, reproducible, rapid and simple means of assessing high-shear platelet function in vitro.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Platelet Function Tests
/
Blood Platelets
/
Hematologic Diseases
/
Hemostasis
Limits:
Humans
Language:
En
Journal:
Blood Coagul Fibrinolysis
Journal subject:
ANGIOLOGIA
/
HEMATOLOGIA
Year:
1999
Document type:
Article
Affiliation country:
United kingdom