Real-time interactive coronary MRA.
Nayak, K S; Pauly, J M; Yang, P C; Hu, B S; Meyer, C H; Nishimura, D G.
Affiliation
  • Nayak KS; Magnetic Resonance Systems Research Laboratory, Department of Electrical Engineering, Stanford University, Stanford, California 94305-9510, USA. nayak@mrsrl.stanford.edu
Magn Reson Med ; 46(3): 430-5, 2001 Sep.
Article in En | MEDLINE | ID: mdl-11550232
An interactive real-time imaging system capable of rapid coronary artery imaging is described. High-resolution spiral and circular echo planar trajectories were used to achieve 0.8 x 1.6 mm2 resolution in 135 ms (CEPI) or 1.13 x 1.13 mm2 resolution in 189 ms (spirals), over a 20-cm FOV. Using a sliding window reconstruction, display rates of up to 37 images/sec were achieved. Initial results indicate this technique can perform as a high-quality 2D coronary localizer and with SNR improvement may enable rapid screening of the coronary tree.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Image Enhancement / Echo-Planar Imaging / Magnetic Resonance Angiography / CD-I Type of study: Diagnostic_studies Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2001 Document type: Article Affiliation country: United States Country of publication: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Image Enhancement / Echo-Planar Imaging / Magnetic Resonance Angiography / CD-I Type of study: Diagnostic_studies Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2001 Document type: Article Affiliation country: United States Country of publication: United States