A rotational approach to localized SPAMM 1-1 tagging.
J Magn Reson
; 157(2): 218-22, 2002 Aug.
Article
in En
| MEDLINE
| ID: mdl-12323140
Magnetic resonance tagging usually relies on controlling the phase dispersion of the transverse magnetization component. Phase dispersion is, however, affected by the inherent phase of selective excitation pulses, thus limiting their combination with tagging sequences to the application of refocusable pulses, as in the localized spatial modulation of magnetization (L-SPAMM) technique. In this study, we examine the effect of selective excitation pulses on a L-SPAMM 1-1 sequence, showing that in the case of two identical pulses the phase component is canceled out, and thus preemphasis and refocus gradients are not needed, allowing us to take advantage of a constant gradient throughout the tagging sequence, and also that one might choose nonrefocusable maximum and minimum phase pulses.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Language:
En
Journal:
J Magn Reson
Journal subject:
DIAGNOSTICO POR IMAGEM
Year:
2002
Document type:
Article
Affiliation country:
Greece
Country of publication:
United States