Guided ultrasonic waves in long bones: modelling, experiment and in vivo application.
Physiol Meas
; 23(4): 755-68, 2002 Nov.
Article
in En
| MEDLINE
| ID: mdl-12450274
Existing ultrasound devices for assessing the human tibia are based on detecting the first arriving signal, corresponding to a wave propagating at, or close to, the bulk longitudinal velocity in bone. However, human long bones are effectively irregular hollow tubes and should theoretically support the propagation of more complex guided modes similar to Lamb waves in plates. Guided waves are attractive because they propagate throughout the bone thickness and can potentially yield more information on bone material properties and architecture. In this study, Lamb wave theory and numerical simulations of wave propagation were used to gain insights into the expected behaviour of guided waves in bone. Experimental measurements in acrylic plates, using a prototype low-frequency axial pulse transmission device, confirmed the presence of two distinct propagating waves: the first arriving wave propagating at, or close to, the longitudinal velocity, and a slower second wave whose behaviour was consistent with the lowest order Lamb antisymmetrical (A0) mode. In a pilot study of healthy and osteoporotic subjects, the velocity of the second wave differed significantly between the two groups, whereas the first arriving wave velocity did not, suggesting the former to be a more sensitive indicator of osteoporosis. We conclude that guided wave measurements may offer an enhanced approach to the ultrasonic characterization of long bones.
Search on Google
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Osteoporosis
/
Tibia
/
Ultrasonography
/
Models, Biological
Type of study:
Diagnostic_studies
Limits:
Humans
Language:
En
Journal:
Physiol Meas
Journal subject:
BIOFISICA
/
ENGENHARIA BIOMEDICA
/
FISIOLOGIA
Year:
2002
Document type:
Article
Affiliation country:
Finland
Country of publication:
United kingdom