Multichannel magnetic stimulation system design considering mutual couplings among the stimulation coils.
Han, Byung H; Chun, In K; Lee, Sang C; Lee, Soo Y.
Affiliation
  • Han BH; Korean Electrotechnology Research Institute, Kyungki, South Korea.
IEEE Trans Biomed Eng ; 51(5): 812-7, 2004 May.
Article in En | MEDLINE | ID: mdl-15132507
ABSTRACT
We introduce some simulation and experiment results of the multichannel magnetic stimulator development that has been carried out as an initial attempt to realize a multichannel functional magnetic stimulator. For efficient functional magnetic stimulations, precise spatial localization of stimulation sites without any movements of the stimulation coils is very important. We have found that the mutual coupling effect among the adjacent stimulation coils in the coil array has to be considered in the determination of the charge voltages in some coil array configurations. Experimental results obtained with a 4-channel magnetic stimulator are presented.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Electric Stimulation Therapy / Equipment Failure Analysis / Electric Stimulation / Electrodes / Equipment Design / Magnetics / Models, Theoretical Type of study: Evaluation_studies Language: En Journal: IEEE Trans Biomed Eng Year: 2004 Document type: Article Affiliation country: South Korea
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Electric Stimulation Therapy / Equipment Failure Analysis / Electric Stimulation / Electrodes / Equipment Design / Magnetics / Models, Theoretical Type of study: Evaluation_studies Language: En Journal: IEEE Trans Biomed Eng Year: 2004 Document type: Article Affiliation country: South Korea