Your browser doesn't support javascript.
loading
[Determination of trace Pb in whole blood by Zeeman-GFAAS].
Chen, Xiao-hong; Yu, Xiao-yu.
Affiliation
  • Chen XH; Department of Laboratory Test, Ningbo Municipal Center for Disease Prevention and Control, Ningbo 315010, China.
Guang Pu Xue Yu Guang Pu Fen Xi ; 25(3): 477-8, 2005 Mar.
Article in Zh | MEDLINE | ID: mdl-16013338
ABSTRACT
The determination method for trace Pb in whole blood was studied, which was diluted by 0.5% HNO3. In the measurement, Zeeman-GFAAS was used and the relative condition was discussed. This method shows that the detection limit was 10.0 pg, the relative standard deviation of standards (RSD) was 0.36%-1.8%, the relative standard deviation of samples (RSD) was 1.4%-2.3%, and the recovery rate was 96.0%-101.7%. The concurrent interference was low. This method is simple and rapid. It is an ideal method to determine trace Pb in whole blood.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Spectrophotometry, Atomic / Graphite / Hot Temperature / Lead Limits: Humans Language: Zh Journal: Guang Pu Xue Yu Guang Pu Fen Xi Year: 2005 Document type: Article Affiliation country: China
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Spectrophotometry, Atomic / Graphite / Hot Temperature / Lead Limits: Humans Language: Zh Journal: Guang Pu Xue Yu Guang Pu Fen Xi Year: 2005 Document type: Article Affiliation country: China