Long-term results of endovascular abdominal aortic aneurysm treatment with the first generation of commercially available stent grafts.
Leurs, Lina J; Buth, Jacob; Laheij, Robert J F.
Affiliation
  • Leurs LJ; EUROSTAR Data Registry Centre, Department of Vascular Surgery, Catharina Hospital, Michelangelolaan 2, 5602 ZA Eindhoven, the Netherlands. EUROstar@iae.nl
Arch Surg ; 142(1): 33-41; discussion 42, 2007 Jan.
Article in En | MEDLINE | ID: mdl-17224498
HYPOTHESIS: Little information about the long-term results of endovascular abdominal aortic aneurysm repair is available. This study was performed to evaluate the long-term data of patients treated with the first generation of commercially available stent grafts. DESIGN: Multicenter registry. SETTING: Sixty-two European centers that participated in the EUROSTAR (EUROpean collaborators on Stent-graft Techniques for abdominal aortic Aneurysm Repair) registry. PATIENTS: A total of 1190 patients with a follow-up of up to 8 years, who underwent endovascular abdominal aortic aneurysm repair with a stent graft (Stentor or Vanguard). INTERVENTION: Elective endovascular abdominal aortic aneurysm repair. MAIN OUTCOME MEASURES: The morbidity and mortality data of patients treated with the first-generation stent graft who enrolled in the EUROSTAR registry were analyzed. Incidence rates of complications were calculated to quantify annual risks. Life-table analyses and multivariate Cox proportional hazards models were used for the survival analysis. RESULTS: Conversion to open repair, aneurysm rupture, all-cause death, and aneurysm-related death occurred in 7.1%, 2.4%, 19.9%, and 3.0% of the patients, respectively. The cumulative percentage of the combined outcome event, conversion-free and rupture-free survival, after 8 years was 48.0%. Procedure-related complications that frequently occurred were endoleak (13.0 cases per 100 patient-years), stenosis/thrombosis (4.6 cases per 100 patient-years), and stent migration (4.3 cases per 100 patient-years). CONCLUSIONS: Patients treated with the first generation of stent grafts will need lifelong surveillance because of a considerable risk of late complications. How these findings translate to the outcome of newer-generation stent grafts is unknown. For this reason, vigilant surveillance remains indicated in all patients who undergo endovascular abdominal aortic aneurysm repair.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Stents / Aortic Aneurysm, Abdominal / Blood Vessel Prosthesis Implantation Type of study: Clinical_trials / Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: Arch Surg Year: 2007 Document type: Article Affiliation country: Netherlands Country of publication: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Stents / Aortic Aneurysm, Abdominal / Blood Vessel Prosthesis Implantation Type of study: Clinical_trials / Prognostic_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: Arch Surg Year: 2007 Document type: Article Affiliation country: Netherlands Country of publication: United States