Your browser doesn't support javascript.
loading
Experimental and Computer-Based Performance Analysis of Two Elastomer VAD Valve Designs.
Escobedo, C; Tovar, F; Suarez, B; Hernandez, A; Corona, F; Sacristan, E.
Affiliation
  • Escobedo C; Innovamédica S.A. de C.V., México.
Conf Proc IEEE Eng Med Biol Soc ; 2005: 7620-3, 2005.
Article in En | MEDLINE | ID: mdl-17282045
The development of a new generation pneumatic Ventricular Assist Device (VAD) required the design of valves for the optimization of its performance. Experiments and computer-based simulations under hydrostatic conditions were analyzed in order to test and compare two low-cost elastomer valve designs. The trileaflet valve design showed a superior hydrostatic performance, having almost a ratio of 1:2 hydraulic resistance than the bileaflet valve design in agreement with both, the experimental and the simulation evidences. This study will address the use of a trileaflet valve designs in the future VAD redesign.
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Conf Proc IEEE Eng Med Biol Soc Journal subject: ENGENHARIA BIOMEDICA Year: 2005 Document type: Article Country of publication: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Conf Proc IEEE Eng Med Biol Soc Journal subject: ENGENHARIA BIOMEDICA Year: 2005 Document type: Article Country of publication: United States