Efficacy of a polyphosphazene nanocoat in reducing thrombogenicity, in-stent stenosis, and inflammatory response in porcine renal and iliac artery stents.
Henn, Christina; Satzl, Stefanie; Christoph, Patricia; Kurz, Patrick; Radeleff, Boris; Stampfl, Ulrike; Stampfl, Sibylle; Berger, Irina; Richter, Goetz Martin.
Affiliation
  • Henn C; Department of Diagnostic Radiology, University Hospital of Heidelberg, Im Neuenheimer Feld 110, 69120 Heidelberg, Germany.
J Vasc Interv Radiol ; 19(3): 427-37, 2008 Mar.
Article in En | MEDLINE | ID: mdl-18295704
ABSTRACT

PURPOSE:

It is hypothesized that the inorganic polymer poly[bis(trifluoroethoxy)phosphazene] (PTFEP) reduces stent-associated thrombosis, in-stent stenosis, and inflammatory response and stimulates reendothelialization in a porcine model. MATERIALS AND

METHODS:

PTFEP-nanocoated and bare stainless-steel stents (316 L) were implanted bilaterally in renal and iliac arteries of 16 minipigs (1, 4, and 12 weeks follow-up durations). Primary study endpoints were thrombogenicity determined by filling defects on angiography and in-stent stenosis assessed by lumen loss on quantitative angiography (ie, percentage stenosis) and light microscopy (ie, neointimal thickness, neointimal area, area stenosis). Secondary endpoints were inflammatory response and reendothelialization evaluated by light microscopy and scanning electron microscopy (SEM), respectively.

RESULTS:

Stent placement was successful in 32 renal and 26 iliac arteries. At follow-up, there were no thrombus depositions on PTFEP-coated renal and iliac stents. Thrombus depositions were found on two of six bare metal iliac stents after 4 weeks (P = .0651). PTFEP-coated stents showed a trend toward reduced in-stent stenosis in all renal and iliac arteries at all intervals in iliac stents, neointimal area was significantly smaller in coated stents than in bare metal stents after 1 week and 4 weeks (P = .03 and P = .001, respectively). In renal arteries, inflammation scores indicated lower inflammatory response in PTFEP-coated stents than in bare metal stents after 1 week (P = .01). After 1 week, coated and bare metal stents exhibited complete reendothelialization on SEM.

CONCLUSIONS:

PTFEP-coated stents exhibited reduced thrombus deposition and a trend toward less in-stent stenosis and inflammatory response than bare metal stents.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Organophosphorus Compounds / Polymers / Thrombosis / Stents / Inflammation Type of study: Prognostic_studies Limits: Animals Language: En Journal: J Vasc Interv Radiol Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2008 Document type: Article Affiliation country: Germany

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Organophosphorus Compounds / Polymers / Thrombosis / Stents / Inflammation Type of study: Prognostic_studies Limits: Animals Language: En Journal: J Vasc Interv Radiol Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2008 Document type: Article Affiliation country: Germany