Pitfalls and prevention strategies for liquid chromatography-tandem mass spectrometry in the selected reaction-monitoring mode for drug analysis.
Sauvage, François-Ludovic; Gaulier, Jean-Michel; Lachâtre, Gérard; Marquet, Pierre.
Affiliation
  • Sauvage FL; CHU Limoges, Department of Pharmacology-Toxicology, Limoges, France.
Clin Chem ; 54(9): 1519-27, 2008 Sep.
Article in En | MEDLINE | ID: mdl-18635749
ABSTRACT

BACKGROUND:

We observed cases of false-positive results with the use of liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS). Different LC-MS/MS techniques that use the selected reaction-monitoring mode, routinely employed for the analysis and quantification of drugs and toxic compounds in biological matrices, were involved in the false-positive and potentially false-positive results obtained. We sought to analyze the causes of and solutions to this problem.

METHODS:

We used a previously reported LC-MS/MS general unknown screening method, as well as manual spectral investigation in 1 case, to perform verification and identification of interfering compounds.

RESULTS:

We observed that false-positive results involved a metabolite of zolpidem that might have been mistaken for lysergic acid diethylamide, benzoylecgonine mistaken for atropine, and clomipramine and 3 phenothiazines that share several common ion transitions.

CONCLUSIONS:

To prevent problems such as those we experienced, we recommend the use of stable-isotope internal standards when possible, relative retention times, 2 transitions or more per compound when possible, and acceptable relative abundance ratios between transitions, with an experience-based tolerance of +/-15% for transitions with a relative abundance >10% and with an extension to +/-25% for transitions <10% when the concentration is at the limit of quantification. A powerful general unknown screening procedure can help to confirm suspected interferences. Our results indicate that the specificity of screening procedures is questionable for LC-MS/MS analyses performed in the selected reaction-monitoring mode and involving a large number of compounds with only 1 transition per compound.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pharmaceutical Preparations / Chromatography, Liquid / Tandem Mass Spectrometry Limits: Humans Language: En Journal: Clin Chem Journal subject: QUIMICA CLINICA Year: 2008 Document type: Article Affiliation country: France

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pharmaceutical Preparations / Chromatography, Liquid / Tandem Mass Spectrometry Limits: Humans Language: En Journal: Clin Chem Journal subject: QUIMICA CLINICA Year: 2008 Document type: Article Affiliation country: France