BACKGROUND: Investigations into the MRI compatibility of middle ear implants made from titanium alloys with 1.5 and 3.0 tesla MRI systems which are frequently used for imaging diagnostics. METHOD: 17 different middle ear (ME) implants (ossicular replacement prosthesis (ORP) and ventilation tubes) made from titanium were tested in vitro. Potential warming was determined via an MRI-compatible fibre optic temperature sensor under the influence of sequences with high-level high frequency impulses. An assessment of the attractive force of the implants was carried out placed on a Petri dish under vibration and floating on rubber-sponge (RS) in a water bath. RESULTS: No significant warming of the implants was observed with any of the used sequences at either 3 or 1.5 Tesla field strength (TF). With 3 TF, all 17 implants changed their position on the surface of the water and moved at a slow speed (0.0004-0.0014 m/s) towards the magnetic field. With 1.5 TF, the tested ME implants moved at a maximum speed of 0.0002 m/s and in the case of the ventilation tubes at 0.0005 m/s. CONCLUSION: No warming occurred in any of the tested middle ear implants at either 1.5 or 3 TF. The attractive forces exerted through the static magnetic field were overall low at 1.5 and 3 TF, indicating that no dislocation is to be expected if intraoperative anchoring is correctly conducted. Nevertheless, the indication for examination at 3 TF should be carefully considered due to the anatomically sensitive region.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Titanium / Ossicular Prosthesis / Magnetic Resonance Imaging / Middle Ear Ventilation Limits: Humans Language: De Journal: Laryngorhinootologie Journal subject: OTORRINOLARINGOLOGIA Year: 2009 Document type: Article Country of publication: Germany

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Titanium / Ossicular Prosthesis / Magnetic Resonance Imaging / Middle Ear Ventilation Limits: Humans Language: De Journal: Laryngorhinootologie Journal subject: OTORRINOLARINGOLOGIA Year: 2009 Document type: Article Country of publication: Germany