Comparison of conventional and sensor-based electronic stethoscopes in detecting cardiac murmurs of dogs.
Vörös, K; Bonnevie, A; Reiczigel, J.
Affiliation
  • Vörös K; Department and Clinic Internal Medicine, Szent István University, Budapest, Hungary. Voros.Karoly@aotk.szie.hu
Tierarztl Prax Ausg K Kleintiere Heimtiere ; 40(2): 103-11, 2012 Apr 24.
Article in En | MEDLINE | ID: mdl-22526814
ABSTRACT
BACKGROUND AND

OBJECTIVE:

Cardiac auscultation is one of the most important parts of the cardiological examination traditionally performed with acoustic stethoscopes. The aim of this study was to compare the sensitivities and the diagnostic capabilities of traditional and electronic stethoscopes in detecting canine heart murmurs. MATERIALS AND

METHODS:

The study was performed on 21 dogs referred for cardiologic examination with suspected heart murmurs. Six out of these dogs had cardiac murmurs bilaterally. Cardiac auscultation was performed independently by a final-year veterinary student (AB=I1) and by an experienced clinician (KV=I2), both using a traditional and a Welch Allyn Meditron electronic sensor-based stethoscope. Final diagnoses were established by echocardiography and by digital phonocardiography.

RESULTS:

Correct detection of a murmur was made by I1 with a traditional stethoscope in 20/27 (74.0%) of the suspected murmurs (p=0.30, kappa[κ] =0.2) and with the electronic stethoscope in 26/27 (96.3%), respectively (p=0.0013, κ=0.75). I2 correctly detected the murmurs with the traditional stethoscope in 25/27 (92.6%) cases (p=0.0013, κ=0.75) and with the electronic stethoscope in all 27/27 (100%) cases (p=0.00012, κ=1). Agreements of murmur intensity gradings between traditional and electronic stethoscopes were highly significant (I1 p=6.9´10⁻8; κ=0.79), (I2 p=5.2´10⁻¹¹; κ=0.92). When grading the murmurs with the traditional stethoscope, there was a significant agreement between I1 and I2 (p=2.9´10⁻7; κ=0.79), being even higher with the electronic stethoscope (p=1.1´10⁻¹¹; κ=0.92).

CONCLUSION:

The electronic stethoscope was more sensitive than the traditional one in detecting and grading cardiac murmurs being especially useful for I1 with less experience. However, it can be suggested to use a traditional and an electronic stethoscopes simultaneously to optimally utilize their advantages.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Heart Murmurs / Stethoscopes / Dog Diseases / Heart Auscultation Type of study: Diagnostic_studies Limits: Animals Language: En Journal: Tierarztl Prax Ausg K Kleintiere Heimtiere Journal subject: MEDICINA VETERINARIA Year: 2012 Document type: Article Affiliation country: Hungary
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Heart Murmurs / Stethoscopes / Dog Diseases / Heart Auscultation Type of study: Diagnostic_studies Limits: Animals Language: En Journal: Tierarztl Prax Ausg K Kleintiere Heimtiere Journal subject: MEDICINA VETERINARIA Year: 2012 Document type: Article Affiliation country: Hungary