On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA.
Med Phys
; 39(6): 3051-9, 2012 Jun.
Article
in En
| MEDLINE
| ID: mdl-22755690
PURPOSE: To investigate using 3D γ analysis for IMRT and VMAT QA. METHODS: We explored and studied 3D γ-analysis by comparing TPS computed and EPID back-projection reconstructed doses in patient's CT images. Two 3D γ quantities, γ(PTV) and γ(10), were proposed and studied for evaluating the QA results, and compared to 2D γ (MapCheck composite: γ(MC)). RESULTS: It was found that when 3%(global)/3 mm criteria was used, all IMRT and 90% of VMAT plans passed QA with a γ pass rate ≥90%. A significant statistical correlation was observed between 3D and 2D γ-analysis results for IMRT QA if γ(10) and γ(MC) are concerned, but no significant relation is found between γ(PTV) and γ(MC). CONCLUSIONS: 3D γ analysis based on EPID dose back-projection may provide a feasible tool for IMRT and VMAT pretreatment plan QA.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Radiotherapy Planning, Computer-Assisted
/
Imaging, Three-Dimensional
/
Radiotherapy, Intensity-Modulated
/
Gamma Rays
Limits:
Humans
/
Male
Language:
En
Journal:
Med Phys
Year:
2012
Document type:
Article
Affiliation country:
United States
Country of publication:
United States