Your browser doesn't support javascript.
loading
Influence of high temperature on solid state nuclear track detector parameters.
Malinowska, A; Szydlowski, A; Jaskóla, M; Korman, A.
Affiliation
  • Malinowska A; National Centre for Nuclear Research (NCBJ), Otwock 05-400, Poland. a.malinowska@ncbj.gov.pl
Rev Sci Instrum ; 83(9): 093502, 2012 Sep.
Article in En | MEDLINE | ID: mdl-23020371
ABSTRACT
This work concerns the influence of high temperatures on tracks induced in solid state nuclear track detectors of the CR-39/PM-355 type. In order to investigate this effect some samples of the detectors were irradiated with energetic protons and α particles and subsequently heated under controlled temperatures for different periods of time. After heating the samples were etched and the track evolution was analyzed using an optical microscope. The bulk etch rate V(B) of the PM-355 material was also determined as a function of heating temperature. The track etch rate V(T) values were estimated for craters induced by protons and α particles from track diameter measurement as a function of heating temperature.
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Rev Sci Instrum Year: 2012 Document type: Article Affiliation country: Poland
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Rev Sci Instrum Year: 2012 Document type: Article Affiliation country: Poland