Your browser doesn't support javascript.
loading
Characterization of the impact of catheter-tissue contact force in lesion formation during cavo-tricuspid isthmus ablation in an experimental swine model.
Matía Francés, Roberto; Hernández Madrid, Antonio; Delgado, Antonia; Carrizo, Laura; Pindado, Carlos; Moro Serrano, Concepción; Zamorano Gómez, José L.
Affiliation
  • Matía Francés R; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain robmatia@gmail.com.
  • Hernández Madrid A; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
  • Delgado A; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
  • Carrizo L; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
  • Pindado C; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
  • Moro Serrano C; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
  • Zamorano Gómez JL; Cardiology Department, Ramón y Cajal Hospital, Ctra. Colmenar Km. 9,1, 28034 Madrid, Spain.
Europace ; 16(11): 1679-83, 2014 Nov.
Article in En | MEDLINE | ID: mdl-24225068
AIMS: Catheter-tissue contact is critical for effective lesion creation. The objective of this study was to determine in an experimental swine model the pathological effects of cavo-tricuspid isthmus ablation using two systems that provide reliable measures of the pressure at the catheter tip during radiofrequency ablation procedures. METHODS AND RESULTS: We performed the procedure in eight pigs in our experimental electrophysiology laboratory after right femoral vein dissection and insertion of a 12 Fr. introducer during general anaesthesia and endotracheal intubation. The target contact force during the applications was <10 grs. (axial or lateral), 10-20, 20-30, and >30 grs. in two pigs each. The power was set at 40 W and maximum target temperature at 45°C. We performed a radiofrequency line dragging from the tricuspid valve to the inferior vena cava in the eight pigs. Euthanasia of the animals was carried out a week after the procedure and a pathological examination of the lesions was performed. In the endocardial macroscopic analysis the extent of lesions, presence of thrombus, transmurality, and endothelial rupture was assessed. External surface was examined searching for transmural lesions. The mean contact force applied was 18.7 ± 8.4 grs. and the mean depth of the lesions was 3.6 ± 2 mm. Lesions were never transmural with average forces <10 grs., and the mean depth was very low (0.75 mm). To achieve transmural lesions contact forces of at least 20 grs. were required. We found a positive correlation (r = 0.85, P < 0.05) between the average force during the applications and depth of the lesions. CONCLUSION: When ablating the cavo-tricuspid isthmus in a swine model, contact forces of at least 20 grs. are required to achieve transmural lesions. Catheter-tissue contact is critical for effective lesion creation. This information is important for improving ablation efficacy.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tricuspid Valve / Vena Cava, Inferior / Catheter Ablation / Cardiac Catheters Type of study: Prognostic_studies Limits: Animals Language: En Journal: Europace Journal subject: CARDIOLOGIA / FISIOLOGIA Year: 2014 Document type: Article Affiliation country: Spain Country of publication: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tricuspid Valve / Vena Cava, Inferior / Catheter Ablation / Cardiac Catheters Type of study: Prognostic_studies Limits: Animals Language: En Journal: Europace Journal subject: CARDIOLOGIA / FISIOLOGIA Year: 2014 Document type: Article Affiliation country: Spain Country of publication: United kingdom