Dynamic inversion time for improved 3D late gadolinium enhancement imaging in patients with atrial fibrillation.
Keegan, Jennifer; Gatehouse, Peter D; Haldar, Shouvik; Wage, Ricardo; Babu-Narayan, Sonya V; Firmin, David N.
Affiliation
  • Keegan J; Cardiovascular Biomedical Research Unit, Royal Brompton and Harefield NHS Trust, London, United Kingdom.
Magn Reson Med ; 73(2): 646-54, 2015 Feb.
Article in En | MEDLINE | ID: mdl-24604664
ABSTRACT

PURPOSE:

High resolution three-dimensional (3D) late gadolinium enhancement (LGE) imaging is performed with single R-wave gating to minimize lengthy acquisition durations. In patients with atrial fibrillation (AF), heart rate variability results in variable magnetization recovery between sequence repeats, and image quality is often poor. In this study, we implemented and tested a dynamic inversion time (dynamic-TI) scheme designed to reduce sequence sensitivity to heart rate variations.

METHODS:

An inversion-prepared 3D segmented gradient echo sequence was modified so that the TI varied automatically from beat-to-beat (dynamic-TI) based on the time since the last sequence repeat. 3D LGE acquisitions were performed in 17 patients prior to radio frequency ablation of persistent AF both with and without dynamic-TI. Qualitative image quality scores, blood signal-to-ghosting ratios (SGRs). and blood-myocardium contrast-to-ghosting ratios (CGRs) were compared.

RESULTS:

Image quality scores were higher with dynamic-TI than without dynamic-TI (2.2 ± 0.9 vs. 1.8 ± 1.1, P = 0.008), as were blood-myocardium CGRs (13.8 ± 7.6 vs. 8.3 ± 6.1, P = 0.003) and blood SGRs (19.6 ± 8.5 vs. 13.1 ± 8.0, P = 0.003).

CONCLUSION:

The dynamic-TI algorithm improves image quality of 3D LGE imaging in this difficult patient population by reducing the sequence sensitivity to RR interval variations
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Organometallic Compounds / Atrial Fibrillation / Algorithms / Image Interpretation, Computer-Assisted / Image Enhancement / Imaging, Three-Dimensional Type of study: Clinical_trials / Diagnostic_studies / Qualitative_research Limits: Aged / Female / Humans / Male / Middle aged Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2015 Document type: Article Affiliation country: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Organometallic Compounds / Atrial Fibrillation / Algorithms / Image Interpretation, Computer-Assisted / Image Enhancement / Imaging, Three-Dimensional Type of study: Clinical_trials / Diagnostic_studies / Qualitative_research Limits: Aged / Female / Humans / Male / Middle aged Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2015 Document type: Article Affiliation country: United kingdom