Your browser doesn't support javascript.
loading
Structure-function correlations in glaucoma using matrix and standard automated perimetry versus time-domain and spectral-domain OCT devices.
Pinto, Luciano Moreira; Costa, Elaine Fiod; Melo, Luiz Alberto S; Gross, Paula Blasco; Sato, Eduardo Toshio; Almeida, Andrea Pereira; Maia, Andre; Paranhos, Augusto.
Affiliation
  • Pinto LM; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Costa EF; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Melo LA; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Gross PB; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Sato ET; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Almeida AP; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Maia A; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
  • Paranhos A; Department of Ophthalmology, Universidade Federal de São Paulo, São Paulo, Brazil.
Invest Ophthalmol Vis Sci ; 55(5): 3074-80, 2014 Apr 10.
Article in En | MEDLINE | ID: mdl-24722699
ABSTRACT

PURPOSE:

We examined the structure-function relationship between two perimetric tests, the frequency doubling technology (FDT) matrix and standard automated perimetry (SAP), and two optical coherence tomography (OCT) devices (time-domain and spectral-domain).

METHODS:

This cross-sectional study included 97 eyes from 29 healthy individuals, and 68 individuals with early, moderate, or advanced primary open-angle glaucoma. The correlations between overall and sectorial parameters of retinal nerve fiber layer thickness (RNFL) measured with Stratus and Spectralis OCT, and the visual field sensitivity obtained with FDT matrix and SAP were assessed. The relationship also was evaluated using a previously described linear model.

RESULTS:

The correlation coefficients for the threshold sensitivity measured with SAP and Stratus OCT ranged from 0.44 to 0.79, and those for Spectralis OCT ranged from 0.30 to 0.75. Regarding FDT matrix, the correlation ranged from 0.40 to 0.79 with Stratus OCT and from 0.39 to 0.79 with Spectralis OCT. Stronger correlations were found in the overall measurements and the arcuate sectors for both visual fields and OCT devices. A linear relationship was observed between FDT matrix sensitivity and the OCT devices. The previously described linear model fit the data from SAP and the OCT devices well, particularly in the inferotemporal sector.

CONCLUSIONS:

The FDT matrix and SAP visual sensitivities were related strongly to the RNFL thickness measured with the Stratus and Spectralis OCT devices, particularly in the overall and arcuate sectors.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Visual Fields / Glaucoma, Open-Angle / Tomography, Optical Coherence / Visual Field Tests / Nerve Fibers Type of study: Observational_studies / Prevalence_studies / Risk_factors_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: Invest Ophthalmol Vis Sci Year: 2014 Document type: Article Affiliation country: Brazil

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Visual Fields / Glaucoma, Open-Angle / Tomography, Optical Coherence / Visual Field Tests / Nerve Fibers Type of study: Observational_studies / Prevalence_studies / Risk_factors_studies Limits: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Language: En Journal: Invest Ophthalmol Vis Sci Year: 2014 Document type: Article Affiliation country: Brazil