The effect of different humeral prosthesis fin designs on shoulder stability: a computational model.
Med Eng Phys
; 36(11): 1382-7, 2014 Nov.
Article
in En
| MEDLINE
| ID: mdl-24736019
ABSTRACT
Humeral prostheses commonly use a fin structure as an attachment point for the supraspinatus muscle in total shoulder arthroplasty (TSA), but these fins may cause injury to the muscle during implantation, inadvertently influencing stability. In order to prevent supraspinatus injury, the effect of different humeral prostheses on shoulder joint stability needs to be investigated. A commercially available prosthesis and two modified humeral prostheses that substituted the fin structure for 2 (2H) or 3 holes (3H) were evaluated using computational models. Glenohumeral abduction was simulated and the superioinferior/anterioposterior stability of the shoulder joint after TSA was calculated. The results revealed that the 2H design had better superioinferior stability than the other prostheses, but was still less stable than the intact shoulder. There were no obvious differences in anterioposterior stability, but the motion patterns were clearly distinguishable from the intact shoulder model. In conclusion, the 2H design showed better superioinferior stability than the 3H design and the commercial product during glenohumeral joint abduction; the three prostheses show similar results in anterioposterior stability. However, the stability of each tested prosthesis was not comparable to the intact shoulder. Therefore, as a compromise, the 2H design should be considered for TSA because of its superior stability.
Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Prostheses and Implants
/
Prosthesis Design
/
Shoulder
/
Computer Simulation
/
Humerus
Limits:
Aged
/
Female
/
Humans
Language:
En
Journal:
Med Eng Phys
Journal subject:
BIOFISICA
/
ENGENHARIA BIOMEDICA
Year:
2014
Document type:
Article
Affiliation country:
Taiwan