Note: deep ultraviolet Raman spectrograph with the laser excitation line down to 177.3 nm and its application.
Rev Sci Instrum
; 85(4): 046105, 2014 Apr.
Article
in En
| MEDLINE
| ID: mdl-24784683
Deep UV Raman spectrograph with the laser excitation line down to 177.3 nm was developed in this laboratory. An ellipsoidal mirror and a dispersed-subtractive triple monochromator were used to collect and disperse Raman light, respectively. The triple monochromator was arranged in a triangular configuration with only six mirrors used. 177.3 nm laser excited Raman spectrum with cut-off wavenumber down to 200 cm(-1) and spectral resolution of 8.0 cm(-1) can be obtained under the condition of high purity N2 purging. With the C-C σ bond in Teflon selectively excited by the 177.3 nm laser, resonance Raman spectrum of Teflon with good quality was recorded on the home-built instrument and the σ-σ(*) transition of C-C bond was studied. The result demonstrates that deep UV Raman spectrograph is powerful for studying the systems with electronic transition located in the deep UV region.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Spectrum Analysis, Raman
/
Ultraviolet Rays
/
Lasers
Language:
En
Journal:
Rev Sci Instrum
Year:
2014
Document type:
Article
Affiliation country:
China
Country of publication:
United States