Drug-eluting balloon versus standard balloon angioplasty for infrapopliteal arterial revascularization in critical limb ischemia: 12-month results from the IN.PACT DEEP randomized trial.
J Am Coll Cardiol
; 64(15): 1568-76, 2014 Oct 14.
Article
in En
| MEDLINE
| ID: mdl-25301459
ABSTRACT
BACKGROUND:
Drug-eluting balloons (DEB) may reduce infrapopliteal restenosis and reintervention rates versus percutaneous transluminal angioplasty (PTA) and improve wound healing/limb preservation.OBJECTIVES:
The goal of this clinical trial was to assess the efficacy and safety of IN.PACT Amphirion drug-eluting balloons (IA-DEB) compared to PTA for infrapopliteal arterial revascularization in patients with critical limb ischemia (CLI).METHODS:
Within a prospective, multicenter, randomized, controlled trial with independent clinical event adjudication and angiographic and wound core laboratories 358 CLI patients were randomized 21 to IA-DEB or PTA. The 2 coprimary efficacy endpoints through 12 months were clinically driven target lesion revascularization (CD-TLR) and late lumen loss (LLL). The primary safety endpoint through 6 months was a composite of all-cause mortality, major amputation, and CD-TLR.RESULTS:
Clinical characteristics were similar between the 2 groups. Significant baseline differences between the IA-DEB and PTA arms included mean lesion length (10.2 cm vs. 12.9 cm; p = 0.002), impaired inflow (40.7% vs. 28.8%; p = 0.035), and previous target limb revascularization (32.2% vs. 21.8%; p = 0.047). Primary efficacy results of IA-DEB versus PTA were CD-TLR of 9.2% versus 13.1% (p = 0.291) and LLL of 0.61 ± 0.78 mm versus 0.62 ± 0.78 mm (p = 0.950). Primary safety endpoints were 17.7% versus 15.8% (p = 0.021) and met the noninferiority hypothesis. A safety signal driven by major amputations through 12 months was observed in the IA-DEB arm versus the PTA arm (8.8% vs. 3.6%; p = 0.080).CONCLUSIONS:
In patients with CLI, IA-DEB had comparable efficacy to PTA. While primary safety was met, there was a trend towards an increased major amputation rate through 12 months compared to PTA. (Study of IN.PACT Amphirion™ Drug Eluting Balloon vs. Standard PTA for the Treatment of Below the Knee Critical Limb Ischemia [INPACT-DEEP]; NCT00941733).Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Popliteal Artery
/
Arterial Occlusive Diseases
/
Vascular Surgical Procedures
/
Angioplasty, Balloon
/
Drug-Eluting Stents
/
Ischemia
/
Leg
Type of study:
Clinical_trials
/
Diagnostic_studies
/
Etiology_studies
/
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Limits:
Aged
/
Female
/
Humans
/
Male
Language:
En
Journal:
J Am Coll Cardiol
Year:
2014
Document type:
Article
Affiliation country:
Germany