Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera.
Gonul, Saban; Koktekir, Bengu Ekinci; Bakbak, Berker; Gedik, Sansal.
Affiliation
  • Gonul S; Department of Ophthalmology, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Koktekir BE; Department of Ophthalmology, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Bakbak B; Department of Ophthalmology, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Gedik S; Department of Ophthalmology, Selcuk University Faculty of Medicine, Konya, Turkey.
Arq Bras Oftalmol ; 77(6): 345-50, 2014.
Article in En | MEDLINE | ID: mdl-25627178
PURPOSE: To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. METHODS: A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs). RESULTS: Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 µm to 28.7 µm for OLCR versus FD-OCT, from -19.2 µm to 30.4 µm for OLCR versus SC, and from -42.6 to 32.3 µm for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98. CONCLUSION: Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cornea / Corneal Topography / Tomography, Optical Coherence / Corneal Pachymetry Limits: Adolescent / Adult / Child / Female / Humans / Male / Middle aged Language: En Journal: Arq Bras Oftalmol Year: 2014 Document type: Article Affiliation country: Turkey Country of publication: Brazil

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cornea / Corneal Topography / Tomography, Optical Coherence / Corneal Pachymetry Limits: Adolescent / Adult / Child / Female / Humans / Male / Middle aged Language: En Journal: Arq Bras Oftalmol Year: 2014 Document type: Article Affiliation country: Turkey Country of publication: Brazil