Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients.
Phys Med Biol
; 60(14): 5527-42, 2015 Jul 21.
Article
in En
| MEDLINE
| ID: mdl-26133986
MR-guided HIFU ablation is a promising technique for the non-invasive treatment of breast cancer. A phase I study was performed to assess the safety and treatment accuracy and precision of MR-HIFU ablation in breast cancer patients (n=10) using a newly developed MR-HIFU platform dedicated to applications in the breast. In this paper a technical analysis of the performance of the dedicated breast MR-HIFU system during breast tumors ablation is described. The main points of investigation were the spatial targeting accuracy and precision of the system and the performance of real-time respiration-corrected MR thermometry.The mean targeting accuracy was in the range of 2.4-2.6 mm, whereas the mean targeting precision was in the range of 1.5-1.8 mm. To correct for respiration-induced magnetic field fluctuations during MR temperature mapping a look-up-table (LUT)-based correction method was used. An optimized procedural sedation protocol in combination with the LUT-based correction method allowed for precise MR thermometry during the ablation procedure (temperature standard deviation <3 °C). No unwanted heating in the near field (i.e. skin) nor in the far field (pectoral muscle) was detected.The newly developed dedicated breast MR-HIFU system allows for safe, accurate and precise ablation of breast tumors.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Image Processing, Computer-Assisted
/
Breast Neoplasms
/
Magnetic Resonance Imaging, Interventional
/
High-Intensity Focused Ultrasound Ablation
/
Thermometry
Type of study:
Guideline
/
Observational_studies
Limits:
Female
/
Humans
Language:
En
Journal:
Phys Med Biol
Year:
2015
Document type:
Article
Affiliation country:
Netherlands
Country of publication:
United kingdom