An improved neutron autoradiography set-up for (10)B concentration measurements in biological samples.
Postuma, Ian; Bortolussi, Silva; Protti, Nicoletta; Ballarini, Francesca; Bruschi, Piero; Ciani, Laura; Ristori, Sandra; Panza, Luigi; Ferrari, Cinzia; Cansolino, Laura; Altieri, Saverio.
Affiliation
  • Postuma I; Department of Physics, University of Pavia, Italy; National Institute of Nuclear Physics, INFN, Section of Pavia, Italy.
  • Bortolussi S; Department of Physics, University of Pavia, Italy; National Institute of Nuclear Physics, INFN, Section of Pavia, Italy.
  • Protti N; Department of Physics, University of Pavia, Italy; National Institute of Nuclear Physics, INFN, Section of Pavia, Italy.
  • Ballarini F; Department of Physics, University of Pavia, Italy; National Institute of Nuclear Physics, INFN, Section of Pavia, Italy.
  • Bruschi P; Department of Physics, University of Pavia, Italy.
  • Ciani L; Department of Chemistry 'UgoSchi' & CSGI, University of Florence, Florence, Italy.
  • Ristori S; Department of Chemistry 'UgoSchi' & CSGI, University of Florence, Florence, Italy.
  • Panza L; DISCAFF, University of Eastern Piedmont, Novara, Italy.
  • Ferrari C; Dipartimento di Scienze Clinico-Chirurgiche, Diagnostiche e Pediatriche, Universit degli Studidi Pavia, Italy.
  • Cansolino L; Dipartimento di Scienze Clinico-Chirurgiche, Diagnostiche e Pediatriche, Universit degli Studidi Pavia, Italy.
  • Altieri S; Department of Physics, University of Pavia, Italy; National Institute of Nuclear Physics, INFN, Section of Pavia, Italy.
Rep Pract Oncol Radiother ; 21(2): 123-8, 2016.
Article in En | MEDLINE | ID: mdl-26933395
ABSTRACT

AIM:

Boron Neutron Capture Therapy (BNCT) is a binary hadrontherapy which exploits the neutron capture reaction in boron, together with a selective uptake of boronated substances by the neoplastic tissue. There is increasing evidence that future improvements in clinical BNCT will be triggered by the discovery of new boronated compounds, with higher selectivity for the tumor with respect to clinically used sodium borocaptate (BSH) and boronophenylalanine (BPA).

BACKGROUND:

Therefore, a (10)B quantification technique for biological samples is needed in order to evaluate the performance of new boronated formulations. MATERIALS AND

METHODS:

This article describes an improved neutron autoradiography set-up employing radiation sensitive films where the latent tracks are made visible by proper etching conditions.

RESULTS:

Calibration curves for both liquid and tissue samples were obtained.

CONCLUSIONS:

The obtained calibration curves were adopted to set-up a mechanism to point out boron concentration in the whole sample.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Rep Pract Oncol Radiother Year: 2016 Document type: Article Affiliation country: Italy Publication country: PL / POLAND / POLONIA / POLÔNIA