Evaluation of an Automated System for Reading and Interpreting Disk Diffusion Antimicrobial Susceptibility Testing of Fastidious Bacteria.
Idelevich, Evgeny A; Becker, Karsten; Schmitz, Janne; Knaack, Dennis; Peters, Georg; Köck, Robin.
Affiliation
  • Idelevich EA; Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
  • Becker K; Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
  • Schmitz J; Institute of Hygiene, University Hospital Münster, Münster, Germany.
  • Knaack D; Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
  • Peters G; Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
  • Köck R; Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
PLoS One ; 11(7): e0159183, 2016.
Article in En | MEDLINE | ID: mdl-27391898
Results of disk diffusion antimicrobial susceptibility testing depend on individual visual reading of inhibition zone diameters. Therefore, automated reading using camera systems might represent a useful tool for standardization. In this study, the ADAGIO automated system (Bio-Rad) was evaluated for reading disk diffusion tests of fastidious bacteria. 144 clinical isolates (68 ß-haemolytic streptococci, 28 Streptococcus pneumoniae, 18 viridans group streptococci, 13 Haemophilus influenzae, 7 Moraxella catarrhalis, and 10 Campylobacter jejuni) were tested on Mueller-Hinton agar supplemented with 5% defibrinated horse blood and 20 mg/L ß-NAD (MH-F, Oxoid) according to EUCAST. Plates were read manually with a ruler and automatically using the ADAGIO system. Inhibition zone diameters, indicated by the automated system, were visually controlled and adjusted, if necessary. Among 1548 isolate-antibiotic combinations, comparison of automated vs. manual reading yielded categorical agreement (CA) without visual adjustment of the automatically determined zone diameters in 81.4%. In 20% (309 of 1548) of tests it was deemed necessary to adjust the automatically determined zone diameter after visual control. After adjustment, CA was 94.8%; very major errors (false susceptible interpretation), major errors (false resistant interpretation) and minor errors (false categorization involving intermediate result), calculated according to the ISO 20776-2 guideline, accounted to 13.7% (13 of 95 resistant results), 3.3% (47 of 1424 susceptible results) and 1.4% (21 of 1548 total results), respectively, compared to manual reading. The ADAGIO system allowed for automated reading of disk diffusion testing in fastidious bacteria and, after visual validation of the automated results, yielded good categorical agreement with manual reading.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Bacteria / Microbial Sensitivity Tests / Anti-Bacterial Agents Type of study: Guideline Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2016 Document type: Article Affiliation country: Germany Country of publication: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Bacteria / Microbial Sensitivity Tests / Anti-Bacterial Agents Type of study: Guideline Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2016 Document type: Article Affiliation country: Germany Country of publication: United States