Combined Fluoroscopic and Arthroscopic Detection and Removal of a Foreign Body Lost During Elective Shoulder Arthroscopy: A Case Report.
Schmiddem, Uli; Hawi, N; Suero, E M; Meller, R.
Affiliation
  • Schmiddem U; Trauma Department, Medical School Hannover, Germany.
  • Hawi N; Trauma Department, Medical School Hannover, Germany.
  • Suero EM; Trauma Department, Medical School Hannover, Germany.
  • Meller R; Trauma Department, Medical School Hannover, Germany.
J Orthop Case Rep ; 7(2): 78-81, 2017.
Article in En | MEDLINE | ID: mdl-28819609
ABSTRACT

INTRODUCTION:

We report a case of a lost metal platelet from a radiofrequency ablation probe (VAPR VUE Radiofrequency System, Cool Pulse 90, DePuy, Synthes, Switzerland) in the shoulder joint during elective arthroscopic cuff repair. To the best of our knowledge, this kind of an incident during elective arthroscopy has not been described in the literature so far. In addition, we present an algorithm on how to deal with such an incident. CASE REPORT A 69-year-old woman underwent an arthroscopic subacromial decompression and rotator cuff repair for a torn supraspinatus tendon. While performing the subacromial decompression and after swapping the portals from lateral to posterior, the metal platelet of the electrocautery device got detached from the instrument and lost in the operation field. Several attempts to visualize the lost platelet with the camera failed. Finally, intraoperative fluoroscopic imaging was used to detect the platelet. To confirm the definitive whereabouts of the platelet, two spinal needles were positioned perpendicular to another under x-ray control, both pointing at the missing platelet. After determining the exact location, the platelet could finally be visualized with the camera and removed. Due to this incident, the operation time was extended extensively, and the patient as well as the theatre team was exposed to an unnecessary amount of radiation.

CONCLUSION:

This report indicates that an extraordinary incident such as the detachment of a component of the arthroscopic equipment during surgery is possible and should be kept in mind by the surgeon. Therefore, we believe that it is essential to perform a test of integrity at least at the end of every operation. In addition, we are presenting an algorithm on how to deal with the situation of a lost foreign body during arthroscopy, which can be applied to any joint.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Type of study: Diagnostic_studies Language: En Journal: J Orthop Case Rep Year: 2017 Document type: Article Affiliation country: Germany Publication country: IN / INDIA / ÍNDIA

Full text: 1 Collection: 01-internacional Database: MEDLINE Type of study: Diagnostic_studies Language: En Journal: J Orthop Case Rep Year: 2017 Document type: Article Affiliation country: Germany Publication country: IN / INDIA / ÍNDIA