Influence of parameter accuracy on pharmacokinetic analysis of hyperpolarized pyruvate.
Sun, Chang-Yu; Walker, Christopher M; Michel, Keith A; Venkatesan, Aradhana M; Lai, Stephen Y; Bankson, James A.
Affiliation
  • Sun CY; Department of Imaging Physics, the University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Walker CM; Department of Imaging Physics, the University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Michel KA; The University of Texas MD Anderson Cancer Center, UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • Venkatesan AM; Department of Imaging Physics, the University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Lai SY; The University of Texas MD Anderson Cancer Center, UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • Bankson JA; Department of Diagnostic Radiology, the University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Magn Reson Med ; 79(6): 3239-3248, 2018 06.
Article in En | MEDLINE | ID: mdl-29090487
ABSTRACT

PURPOSE:

To explore the effects of noise and error on kinetic analyses of tumor metabolism using hyperpolarized [1-13 C] pyruvate.

METHODS:

Numerical simulations were performed to systematically investigate the effects of noise, the number of unknowns, and error in kinetic parameter estimates on kinetic analysis of the apparent rate of chemical conversion from hyperpolarized pyruvate to lactate (kPL ). A pharmacokinetic model with two physical and two chemical pools of hyperpolarized spins was used to generate and analyze the synthetic data.

RESULTS:

The reproducibility of kPL estimates worsened quickly when peak signal-to-noise ratio for hyperpolarized pyruvate was below approximately 20. The accuracy of kPL estimates was most sensitive to errors in high excitation angles, the vascular blood volume fraction (vb ), and the rate of pyruvate extravasation (kve ), and was least sensitive to errors in the T1 of pyruvate. When vb and/or kve were fit as additional unknowns, the accuracy of kPL estimates suffered, and when the vascular input function of pyruvate was also fit, the reproducibility of kPL estimates worsened.

CONCLUSIONS:

The accuracy and precision of kPL estimates improve substantially for peak signal-to-noise ratio above approximately 20. Accurate estimates of perfusion parameters (combinations of vb , kve , and the pyruvate vascular input function) and transmit calibration at high excitation angles have the greatest effect on the accuracy of kinetic analyses. Magn Reson Med 793239-3248, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Carbon Isotopes / Magnetic Resonance Imaging / Pyruvic Acid / Neoplasms Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2018 Document type: Article Affiliation country: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Carbon Isotopes / Magnetic Resonance Imaging / Pyruvic Acid / Neoplasms Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2018 Document type: Article Affiliation country: United States