Your browser doesn't support javascript.
loading
Influence of varus/valgus positioning of the Nanos® and Metha® short-stemmed prostheses on stress shielding of metaphyseal bone.
Acta Orthop Belg ; 83(1): 57-66, 2017 Mar.
Article in En | MEDLINE | ID: mdl-29322896
ABSTRACT
The aim of this study was to analyze bone remodeling around the Nanos® (Smith & Nephew) and Metha® (Aesculap AG) implants as a function of varus/valgus stem positioning. In 75 patients with diagnosed coxarthrosis, either Nanos® (n= 51) or Metha® (n= 24) prostheses were implanted. Digital assessment of plain radiographs immediately, 97 days, and 381 days after THA showed no clinically-relevant migration, angulation, or change in offset and center of rotation. The DEXA scans showed significant BMD changes in Gruen zones 1 (-12.8%), 2 (-3.3%), 6 (+6.4%), and 7(-7.8%)(t-test). The pre/postoperative CCD for the Nanos® was 129°/ 135° and for the Metha® 131°/ 127°. Linear regression analysis showed no prediction for BMD by postoperative CCD or stem type. In conclusion, there was no clinically-relevant influence on proximal femur BMD according to varus/valgus implantation of the Nanos® or Metha® prostheses.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Bone Density / Arthroplasty, Replacement, Hip / Femur / Hip Joint / Hip Prosthesis Type of study: Etiology_studies Limits: Humans Language: En Journal: Acta Orthop Belg Year: 2017 Document type: Article
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Bone Density / Arthroplasty, Replacement, Hip / Femur / Hip Joint / Hip Prosthesis Type of study: Etiology_studies Limits: Humans Language: En Journal: Acta Orthop Belg Year: 2017 Document type: Article