Your browser doesn't support javascript.
loading
Determination of a suitable low-dose abdominopelvic CT protocol using model-based iterative reconstruction through cadaveric study.
Moloney, Fiachra; Twomey, Maria; Fama, Daniel; Balta, Joy Y; James, Karl; Kavanagh, Richard G; Moore, Niamh; Murphy, Mary Jane; O'Mahony, Siobhan M; Maher, Michael M; Cryan, John F; O'Connor, Owen J.
Affiliation
  • Moloney F; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • Twomey M; Department of Anatomy and Neuroscience, College of Medicine and Health, University College Cork, Cork, Ireland.
  • Fama D; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • Balta JY; Department of Anatomy and Neuroscience, College of Medicine and Health, University College Cork, Cork, Ireland.
  • James K; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • Kavanagh RG; Department of Anatomy and Neuroscience, College of Medicine and Health, University College Cork, Cork, Ireland.
  • Moore N; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • Murphy MJ; Department of Anatomy and Neuroscience, College of Medicine and Health, University College Cork, Cork, Ireland.
  • O'Mahony SM; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • Maher MM; Department of Anatomy and Neuroscience, College of Medicine and Health, University College Cork, Cork, Ireland.
  • Cryan JF; Department of Radiology, Cork University Hospital, Cork, Ireland.
  • O'Connor OJ; Department of Radiology, Cork University Hospital, Cork, Ireland.
J Med Imaging Radiat Oncol ; 62(5): 625-633, 2018 Oct.
Article in En | MEDLINE | ID: mdl-29656596
INTRODUCTION: Cadaveric studies provide a means of safely assessing new technologies and optimizing scanning prior to clinical validation. Reducing radiation exposure in a clinical setting can entail incremental dose reductions to avoid missing important clinical findings. The use of cadavers allows assessment of the impact of more substantial dose reductions on image quality. Our aim was to identify a suitable low-dose abdominopelvic CT protocol for subsequent clinical validation. METHODS: Five human cadavers were scanned at one conventional dose and three low-dose settings. All scans were reconstructed using three different reconstruction algorithms: filtered back projection (FBP), hybrid iterative reconstruction (60% FBP and 40% adaptive statistical iterative reconstruction (ASIR40)), and model-based iterative reconstruction (MBIR). Two readers rated the image quality both quantitatively and qualitatively. RESULTS: Model-based iterative reconstruction images had significantly better objective image noise and higher qualitative scores compared with both FBP and ASIR40 images at all dose levels. The greatest absolute noise reduction, between MBIR and FBP, of 34.3 HU (equating to a 68% reduction) was at the lowest dose level. MBIR reduced image noise and improved image quality even in CT images acquired with a mean radiation dose reduction of 62% compared with conventional dose studies reconstructed with ASIR40, with lower levels of objective image noise, superior diagnostic acceptability and contrast resolution, and comparable subjective image noise and streak artefact scores. CONCLUSION: This cadaveric study demonstrates that MBIR reduces image noise and improves image quality in abdominopelvic CT images acquired with dose reductions of up to 62%.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiation Dosage / Radiography, Abdominal / Radiographic Image Interpretation, Computer-Assisted / Tomography, X-Ray Computed Type of study: Guideline / Prognostic_studies / Qualitative_research Limits: Humans Language: En Journal: J Med Imaging Radiat Oncol Journal subject: DIAGNOSTICO POR IMAGEM / NEOPLASIAS / RADIOLOGIA Year: 2018 Document type: Article Affiliation country: Ireland Country of publication: Australia

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiation Dosage / Radiography, Abdominal / Radiographic Image Interpretation, Computer-Assisted / Tomography, X-Ray Computed Type of study: Guideline / Prognostic_studies / Qualitative_research Limits: Humans Language: En Journal: J Med Imaging Radiat Oncol Journal subject: DIAGNOSTICO POR IMAGEM / NEOPLASIAS / RADIOLOGIA Year: 2018 Document type: Article Affiliation country: Ireland Country of publication: Australia