Your browser doesn't support javascript.
loading
Improved sensitivity and limit-of-detection of lateral flow devices using spatial constrictions of the flow-path.
Katis, Ioannis N; He, Peijun J W; Eason, Robert W; Sones, Collin L.
Affiliation
  • Katis IN; Optoelectronics Research Centre, University of Southampton, Highfield Campus, SO17 1BJ Southampton, UK. Electronic address: I.Katis@soton.ac.uk.
  • He PJW; Optoelectronics Research Centre, University of Southampton, Highfield Campus, SO17 1BJ Southampton, UK.
  • Eason RW; Optoelectronics Research Centre, University of Southampton, Highfield Campus, SO17 1BJ Southampton, UK.
  • Sones CL; Optoelectronics Research Centre, University of Southampton, Highfield Campus, SO17 1BJ Southampton, UK.
Biosens Bioelectron ; 113: 95-100, 2018 Aug 15.
Article in En | MEDLINE | ID: mdl-29738945
ABSTRACT
We report on the use of a laser-direct write (LDW) technique that allows the fabrication of lateral flow devices with enhanced sensitivity and limit of detection. This manufacturing technique comprises the dispensing of a liquid photopolymer at specific regions of a nitrocellulose membrane and its subsequent photopolymerisation to create impermeable walls inside the volume of the membrane. These polymerised structures are intentionally designed to create fluidic channels which are constricted over a specific length that spans the test zone within which the sample interacts with pre-deposited reagents. Experiments were conducted to show how these constrictions alter the fluid flow rate and the test zone area within the constricted channel geometries. The slower flow rate and smaller test zone area result in the increased sensitivity and lowered limit of detection for these devices. We have quantified these via the improved performance of a C-Reactive Protein (CRP) sandwich assay on our lateral flow devices with constricted flow paths which demonstrate an improvement in its sensitivity by 62x and in its limit of detection by 30x when compared to a standard lateral flow CRP device.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: C-Reactive Protein / Biosensing Techniques / Collodion / Microfluidic Analytical Techniques / Membranes, Artificial Type of study: Diagnostic_studies Limits: Humans Language: En Journal: Biosens Bioelectron Journal subject: BIOTECNOLOGIA Year: 2018 Document type: Article Publication country: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: C-Reactive Protein / Biosensing Techniques / Collodion / Microfluidic Analytical Techniques / Membranes, Artificial Type of study: Diagnostic_studies Limits: Humans Language: En Journal: Biosens Bioelectron Journal subject: BIOTECNOLOGIA Year: 2018 Document type: Article Publication country: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM