Your browser doesn't support javascript.
loading
Morphometrical and biomechanical analyses of a stentless bioprosthetic valve: an implication to avoid potential primary tissue failure.
Takaya, Hiroki; Masuda, Shinya; Naganuma, Masaaki; Yoshioka, Ichiro; Takahashi, Goro; Akiyama, Masatoshi; Adachi, Osamu; Kumagai, Kiichiro; Sugita, Shukei; Saiki, Yoshikatsu.
Affiliation
  • Takaya H; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan. yoshisaiki@med.tohoku.ac.jp.
  • Masuda S; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Naganuma M; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Yoshioka I; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Takahashi G; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Akiyama M; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Adachi O; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
  • Kumagai K; Research Division of Sciences for Aortic Disease, Tohoku University Graduate School of Medicine, Sendai, Japan.
  • Sugita S; Biomechanics Laboratory, Department of Mechanical Engineering, Graduate School of Engineering, Nagoya Institute of Technology, Nagoya, Japan.
  • Saiki Y; Division of Cardiovascular Surgery, Tohoku University Graduate School of Medicine, 1-1 Seiryocho, Aoba-ku, Sendai, Miyagi, 980-8575, Japan.
Gen Thorac Cardiovasc Surg ; 66(9): 523-528, 2018 Sep.
Article in En | MEDLINE | ID: mdl-29956049
ABSTRACT

OBJECTIVES:

Stentless bioprosthetic valves provide hemodynamic advantages over stented valves as well as excellent durability. However, some primary tissue failures in bioprostheses have been reported. This study was conducted to evaluate the morphometrical and biomechanical properties of the stentless Medtronic Freestyle™ aortic root bioprosthesis, to identify any arising problem areas, and to speculate on a potential solution.

METHODS:

The three-dimensional heterogeneity of the stentless bioprosthesis wall was investigated using computed tomography. The ascending aorta and the right, left, and non-coronary sinuses of Valsalva were resected and examined by an indentation test to evaluate their biomechanical properties.

RESULTS:

The non-coronary sinus of Valsalva was significantly thinner than the right sinus of Valsalva (p < 0.01). Young's modulus, calculated as an indicator of elasticity, was significantly greater at the non-coronary sinus of Valsalva (430.7 ± 374.2 kPa) than at either the left (190.6 ± 70.6 kPa, p < 0.01) or right sinuses of Valsalva (240.0 ± 56.5 kPa, p < 0.05).

CONCLUSIONS:

Based on the morphometrical and biomechanical analyses of the stentless bioprosthesis, we demonstrated that there are differences in wall thickness and elasticity between each sinus of Valsalva. These differences suggest that the non-coronary sinus of Valsalva is the most vulnerable and at greater risk of tissue failure. The exclusion of the non-coronary sinus of Valsalva may be beneficial to mitigate the long-term risks of tissue failure in the stentless bioprosthesis.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Aortic Valve Stenosis / Sinus of Valsalva / Bioprosthesis / Heart Valve Prosthesis / Heart Valve Prosthesis Implantation Limits: Aged / Female / Humans / Male Language: En Journal: Gen Thorac Cardiovasc Surg Journal subject: ANGIOLOGIA / CARDIOLOGIA Year: 2018 Document type: Article Affiliation country: Japan

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Aortic Valve Stenosis / Sinus of Valsalva / Bioprosthesis / Heart Valve Prosthesis / Heart Valve Prosthesis Implantation Limits: Aged / Female / Humans / Male Language: En Journal: Gen Thorac Cardiovasc Surg Journal subject: ANGIOLOGIA / CARDIOLOGIA Year: 2018 Document type: Article Affiliation country: Japan
...