Your browser doesn't support javascript.
loading
Theoretical and experimental analysis of photon counting detector CT for proton stopping power prediction.
Taasti, Vicki T; Hansen, David C; Michalak, Gregory J; Deisher, Amanda J; Kruse, Jon J; Muren, Ludvig P; Petersen, Jørgen B B; McCollough, Cynthia H.
Affiliation
  • Taasti VT; Department of Medical Physics, Aarhus University Hospital, Aarhus, Denmark.
  • Hansen DC; Department of Medical Physics, Aarhus University Hospital, Aarhus, Denmark.
  • Michalak GJ; Department of Radiology, Mayo Clinic, Rochester, MN, USA.
  • Deisher AJ; Department of Radiation Oncology, Mayo Clinic, Rochester, MN, USA.
  • Kruse JJ; Department of Radiation Oncology, Mayo Clinic, Rochester, MN, USA.
  • Muren LP; Department of Medical Physics, Aarhus University Hospital, Aarhus, Denmark.
  • Petersen JBB; Department of Medical Physics, Aarhus University Hospital, Aarhus, Denmark.
  • McCollough CH; Department of Radiology, Mayo Clinic, Rochester, MN, USA.
Med Phys ; 45(11): 5186-5196, 2018 Nov.
Article in En | MEDLINE | ID: mdl-30191573
ABSTRACT

PURPOSE:

Photon counting detectors (PCDs) are being introduced in advanced x-ray computed tomography (CT) scanners. From a single PCD-CT acquisition, multiple images can be reconstructed, each based on only a part of the original x-ray spectrum. In this study, we investigated whether PCD-CT can be used to estimate stopping power ratios (SPRs) for proton therapy treatment planning, both by comparing to other SPR methods proposed for single energy CT (SECT) and dual energy CT (DECT) as well as to experimental measurements.

METHODS:

A previously developed DECT-based SPR estimation method was adapted to PCD-CT data, by adjusting the estimation equations to allow for more energy spectra. The method was calibrated directly on noisy data to increase the robustness toward image noise. The new PCD SPR estimation method was tested in theoretical calculations as well as in an experimental setup, using both four and two energy bin PCD-CT images, and through comparison to two other SPR methods proposed for SECT and DECT. These two methods were also evaluated on PCD-CT images, full spectrum (one-bin) or two-bin images, respectively. In a theoretical framework, we evaluated the effect of patient-specific tissue variations (density and elemental composition) and image noise on the SPR accuracy; the latter effect was assessed by applying three different noise levels (low, medium, and high noise). SPR estimates derived using real PCD-CT images were compared to experimentally measured SPRs in nine organic tissue samples, including fat, muscle, and bone tissues.

RESULTS:

For the theoretical calculations, the root-mean-square error (RMSE) of the SPR estimation was 0.1% for the new PCD method using both two and four energy bins, compared to 0.2% and 0.7% for the DECT- and SECT-based method, respectively. The PCD method was found to be very robust toward CT image noise, with a RMSE of 2.7% when high noise was added to the CT numbers. Introducing tissue variations, the RMSE only increased to 0.5%; even when adding high image noise to the changed tissues, the RMSE stayed within 3.1%. In the experimental measurements, the RMSE over the nine tissue samples was 0.8% when using two energy bins, and 1.0% for the four-bin images.

CONCLUSIONS:

In all tested cases, the new PCD method produced similar or better results than the SECT- and DECT-based methods, showing an overall improvement of the SPR accuracy. This study thus demonstrated that PCD-CT scans will be a qualified candidate for SPR estimations.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Protons / Tomography, X-Ray Computed / Photons Type of study: Prognostic_studies / Risk_factors_studies Language: En Journal: Med Phys Year: 2018 Document type: Article Affiliation country: Denmark

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Protons / Tomography, X-Ray Computed / Photons Type of study: Prognostic_studies / Risk_factors_studies Language: En Journal: Med Phys Year: 2018 Document type: Article Affiliation country: Denmark