Measuring depth of anaesthesia using changes in directional connectivity: a comparison with auditory middle latency response and estimated bispectral index during propofol anaesthesia.
Anaesthesia
; 74(3): 321-332, 2019 Mar.
Article
in En
| MEDLINE
| ID: mdl-30556186
General anaesthesia is associated with changes in connectivity between different regions of the brain, the assessment of which has the potential to provide a novel marker of anaesthetic effect. We propose an index that quantifies the strength and direction of information flow in electroencephalographic signals collected across the scalp, assess its performance in discriminating 'wakefulness' from 'anaesthesia', and compare it with estimated bispectral index and the auditory middle latency response. We used a step-wise slow induction of anaesthesia in 10 patients to assess graded changes in electroencephalographic directional connectivity at propofol effect-site concentrations of 2 µg.ml-1 , 3 µg.ml-1 and 4 µg.ml-1 . For each stable effect-site concentration, connectivity was estimated from multichannel electroencephalograms using directed coherence, together with middle latency response and estimated bispectral index. We used a linear support vector machine classifier to compare the performance of the different electroencephalographic features in discriminating wakefulness from anaesthesia. We found a significant reduction in the strength of long-range connectivity (interelectrode distance > 10 cm) (p < 0.008), and a reversal of information flow from markedly postero-frontal to fronto-posterior (p < 0.006) between wakefulness and a propofol effect-site concentration of 2 µg.ml-1 . This then remained relatively constant as effect-site concentration increased, consistent with a step change in directed coherence with anaesthesia. This contrasted with the gradual change with increasing anaesthetic dose observed for estimated bispectral index and middle latency response. Directed coherence performed best in discriminating wakefulness from anaesthesia with an accuracy of 95%, indicating the potential of this new method (on its own or combined with others) for monitoring adequacy of anaesthesia.
Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Reaction Time
/
Propofol
/
Electroencephalography
/
Anesthesia, Intravenous
Limits:
Adult
/
Aged
/
Humans
/
Middle aged
Language:
En
Journal:
Anaesthesia
Year:
2019
Document type:
Article
Country of publication:
United kingdom