Your browser doesn't support javascript.
loading
Design and Application of Individualized, 3-Dimensional-Printed Navigation Template for Placing Cortical Bone Trajectory Screws in Middle-Upper Thoracic Spine: Cadaver Research Study.
Wang, Ke; Zhang, Zeng-Jie; Chen, Jiao-Xiang; Wu, Ai-Min; Wang, Xiang-Yang; Sheng, Sun-Ren.
Affiliation
  • Wang K; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
  • Zhang ZJ; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
  • Chen JX; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
  • Wu AM; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
  • Wang XY; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China.
  • Sheng SR; Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children's Hospital of Wenzhou Medical University, Wenzhou, China. Electronic address: shengsunren82330@126.com.
World Neurosurg ; 125: e348-e352, 2019 05.
Article in En | MEDLINE | ID: mdl-30703596
ABSTRACT

OBJECTIVE:

To evaluate the safety and accuracy of use of a 3-dimensional printed navigation template in the placement of a cortical bone trajectory (CBT) screw in the middle-upper thoracic spine.

METHODS:

Ten human cadavers were included in the study. Sixty CBT screws were placed on 1 side, using the free-hand technique, and 60 CBT screws were placed on the other side, using the navigation template that was designed and printed using data from 10 cadavers. The safety and accuracy of use of the CBT screws were directly evaluated by radiography and computed tomography.

RESULTS:

Computed tomography revealed that 2 and 3 of 60 screws, placed using the navigation template, were broken in the medial or lateral areas and in the superior or inferior pedicle wall, respectively. Furthermore, 8 screws were broken in the medial or lateral areas and 11 screws were broken in the superior or inferior pedicle wall when the free-hand technique was used. Radiography revealed that 3 screws in zone I, 55 screws in zone II, and 2 screws in zone III were placed using the navigation template. Furthermore, 7 screws in zone I, 45 screws in zone II, and 8 screws in zone III were placed using the free-hand technique.

CONCLUSIONS:

In this cadaver study, insertion of the CBT screws in the middle-upper thoracic spine with the assistance of the navigation template was safe and convenient.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Thoracic Vertebrae / Pedicle Screws / Printing, Three-Dimensional / Cortical Bone Limits: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Language: En Journal: World Neurosurg Journal subject: NEUROCIRURGIA Year: 2019 Document type: Article Affiliation country: China

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Thoracic Vertebrae / Pedicle Screws / Printing, Three-Dimensional / Cortical Bone Limits: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Language: En Journal: World Neurosurg Journal subject: NEUROCIRURGIA Year: 2019 Document type: Article Affiliation country: China