A comparison of static and dynamic ∆B0 mapping methods for correction of CEST MRI in the presence of temporal B0 field variations.
Poblador Rodriguez, Esau; Moser, Philipp; Dymerska, Barbara; Robinson, Simon; Schmitt, Benjamin; van der Kouwe, Andre; Gruber, Stephan; Trattnig, Siegfried; Bogner, Wolfgang.
Affiliation
  • Poblador Rodriguez E; High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
  • Moser P; High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
  • Dymerska B; Medical Physics and Bioengineering, University College London, London, United Kingdom.
  • Robinson S; High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
  • Schmitt B; Siemens Healthineers, Sydney, Australia.
  • van der Kouwe A; Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts.
  • Gruber S; High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
  • Trattnig S; High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University Vienna, Vienna, Austria.
  • Bogner W; Christian Doppler Laboratory for Clinical Molecular MR Imaging, Vienna, Austria.
Magn Reson Med ; 82(2): 633-646, 2019 08.
Article in En | MEDLINE | ID: mdl-30924210
ABSTRACT

PURPOSE:

To assess the performance, in the presence of scanner instabilities, of three dynamic correction methods which integrate ∆B0 mapping into the chemical exchange saturation transfer (CEST) measurement and three established static ∆B0 -correction approaches.

METHODS:

A homogeneous phantom and five healthy volunteers were scanned with a CEST sequence at 7 T. The in vivo measurements were performed twice first with unaltered system frequency and again applying frequency shifts during the CEST acquisition. In all cases, retrospective voxel-wise ∆B0 -correction was performed using one intrinsic and two extrinsic [prescans with dual-echo gradient-echo and water saturation shift referencing (WASSR)] static approaches. These were compared with two intrinsic [using phase data directly generated by single-echo or double-echo GRE (gradient-echo) CEST readout (CEST-GRE-2TE)] and one extrinsic [phase from interleaved dual-echo EPI (echo planar imaging) navigator (NAV-EPI-2TE)] dynamic ∆B0 -correction approaches [allowing correction of each Z-spectral point before magnetization transfer ratio asymmetry (MTRasym) analysis].

RESULTS:

All three dynamic methods successfully mapped the induced drift. The intrinsic approaches were affected by the CEST labeling near water (∆ω < |0.3| ppm). The MTRasym contrast was distorted by the frequency drift in the brain by up to 0.21%/Hz when static ∆B0 -corrections were applied, whereas the dynamic ∆B0 corrections reduced this to <0.01%/Hz without the need of external scans. The CEST-GRE-2TE and NAV-EPI-2TE resulted in highly consistent MTRasym values with/without drift for all subjects.

CONCLUSION:

Reliable correction of scanner instabilities is essential to establish clinical CEST MRI. The three dynamic approaches presented improved the ∆B0 -correction performance significantly in the presence of frequency drift compared to established static methods. Among them, the self-corrected CEST-GRE-2TE was the most accurate and straightforward to implement.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Magnetic Resonance Imaging Limits: Adult / Female / Humans / Male Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2019 Document type: Article Affiliation country: Austria

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Magnetic Resonance Imaging Limits: Adult / Female / Humans / Male Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2019 Document type: Article Affiliation country: Austria