Your browser doesn't support javascript.
loading
Impact of Leadless Pacemaker Therapy on Cardiac and Atrioventricular Valve Function Through 12 Months of Follow-Up.
Beurskens, Niek E G; Tjong, Fleur V Y; de Bruin-Bon, Rianne H A; Dasselaar, Kosse J; Kuijt, Wichert J; Wilde, Arthur A M; Knops, Reinoud E.
Affiliation
  • Beurskens NEG; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • Tjong FVY; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • de Bruin-Bon RHA; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • Dasselaar KJ; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • Kuijt WJ; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • Wilde AAM; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
  • Knops RE; Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam UMC, University of Amsterdam, Cardiovascular Sciences, the Netherlands.
Circ Arrhythm Electrophysiol ; 12(5): e007124, 2019 05.
Article in En | MEDLINE | ID: mdl-31060371
ABSTRACT

BACKGROUND:

Endocardial pacemaker leads and right ventricular (RV) pacing are well-known causes of tricuspid valve, mitral valve, and cardiac dysfunction. Lead-related adverse consequences can potentially be mitigated by leadless pacemaker (LP) therapy by eliminating the presence of a transvalvular lead. This study assessed the impact of LP placement on cardiac and valvular structure and function.

METHODS:

Echocardiographic studies before and 12±1 months after LP implantation were performed between January 2013 and May 2018 at our center and compared with age- and sex-matched controls of dual-chamber transvenous pacemaker recipients.

RESULTS:

A total of 53 patients receiving an LP were included, of whom 28 were implanted with a Nanostim and 25 with a Micra LP device. Tricuspid valve regurgitation was graded as being more severe in 23 (43%) patients at 12±1 months compared with baseline ( P<0.001). Compared with an apical position, an RV septal position of the LP was associated with increased tricuspid valve incompetence (odds ratio, 5.20; P=0.03). An increase in mitral valve regurgitation was observed in 38% of patients ( P=0.006). LP implantation resulted in a reduction of RV function, according to a lower tricuspid annular plane systolic excursion ( P=0.003) and RV tricuspid lateral annular systolic velocity ( P=0.02), and a higher RV Tei index ( P=0.04). LP implantation was further associated with a reduction of left ventricular ejection fraction ( P=0.03) and elevated left ventricular Tei index ( P=0.003). The changes in tricuspid valve regurgitation in the LP group were similar to the changes in the dual-chamber transvenous pacemaker control group (43% versus 38%, respectively; P=0.39).

CONCLUSIONS:

LP therapy is associated with an increase in tricuspid valve dysfunction through 12 months of follow-up; yet it was comparable to dual-chamber transvenous pacemaker systems. Furthermore, LP therapy seems to adversely impact mitral valve and biventricular function.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pacemaker, Artificial / Tricuspid Valve / Tricuspid Valve Insufficiency / Cardiac Pacing, Artificial / Ventricular Function, Left / Ventricular Function, Right / Ventricular Dysfunction, Right / Ventricular Dysfunction, Left / Mitral Valve / Mitral Valve Insufficiency Type of study: Etiology_studies / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male Language: En Journal: Circ Arrhythm Electrophysiol Journal subject: ANGIOLOGIA / CARDIOLOGIA Year: 2019 Document type: Article Affiliation country: Netherlands Publication country: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Pacemaker, Artificial / Tricuspid Valve / Tricuspid Valve Insufficiency / Cardiac Pacing, Artificial / Ventricular Function, Left / Ventricular Function, Right / Ventricular Dysfunction, Right / Ventricular Dysfunction, Left / Mitral Valve / Mitral Valve Insufficiency Type of study: Etiology_studies / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male Language: En Journal: Circ Arrhythm Electrophysiol Journal subject: ANGIOLOGIA / CARDIOLOGIA Year: 2019 Document type: Article Affiliation country: Netherlands Publication country: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA