Utility of a smartphone based system (cvrphone) to accurately determine apneic events from electrocardiographic signals.
Sohn, Kwanghyun; Merchant, Faisal M; Abohashem, Shady; Kulkarni, Kanchan; Singh, Jagmeet P; Heist, E Kevin; Owen, Chris; Roberts, Jesse D; Isselbacher, Eric M; Sana, Furrukh; Armoundas, Antonis A.
Affiliation
  • Sohn K; Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA, United States of America.
  • Merchant FM; Cardiology Division, Emory University School of Medicine, Atlanta, GA, United States of America.
  • Abohashem S; Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA, United States of America.
  • Kulkarni K; Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA, United States of America.
  • Singh JP; Cardiology Division, Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, MA, United States of America.
  • Heist EK; Cardiology Division, Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, MA, United States of America.
  • Owen C; Neurosurgery Division, Massachusetts General Hospital, Boston, MA, United States of America.
  • Roberts JD; Anesthesia, Critical Care and Pain Medicine, Massachusetts General Hospital, Boston, MA, United States of America.
  • Isselbacher EM; Healthcare Transformation Lab, Massachusetts General Hospital, Boston, MA, United States of America.
  • Sana F; Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA, United States of America.
  • Armoundas AA; Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA, United States of America.
PLoS One ; 14(6): e0217217, 2019.
Article in En | MEDLINE | ID: mdl-31206522
ABSTRACT

BACKGROUND:

Sleep disordered breathing manifested as sleep apnea (SA) is prevalent in the general population, and while it is associated with increased morbidity and mortality risk in some patient populations, it remains under-diagnosed. The objective of this study was to assess the accuracy of respiration-rate (RR) and tidal-volume (TV) estimation algorithms, from body-surface ECG signals, using a smartphone based ambulatory respiration monitoring system (cvrPhone).

METHODS:

Twelve lead ECG signals were collected using the cvrPhone from anesthetized and mechanically ventilated swine (n = 9). During ECG data acquisition, the mechanical ventilator tidal-volume (TV) was varied from 250 to 0 to 750 to 0 to 500 to 0 to 750 ml at respiratory rates (RR) of 6 and 14 breaths/min, respectively, and the RR and TV values were estimated from the ECG signals using custom algorithms.

RESULTS:

TV estimations from any two different TV settings showed statistically significant difference (p < 0.01) regardless of the RR. RRs were estimated to be 6.1±1.1 and 14.0±0.2 breaths/min at 6 and 14 breaths/min, respectively (when 250, 500 and 750 ml TV settings were combined). During apnea, the estimated TV and RR values were 11.7±54.9 ml and 0.0±3.5 breaths/min, which were significantly different (p<0.05) than TV and RR values during non-apnea breathing. In addition, the time delay from the apnea onset to the first apnea detection was 8.6±6.7 and 7.0±3.2 seconds for TV and RR respectively.

CONCLUSIONS:

We have demonstrated that apnea can reliably be detected using ECG-derived RR and TV algorithms. These results support the concept that our algorithms can be utilized to detect SA in conjunction with ECG monitoring.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Sleep Apnea Syndromes / Electrocardiography / Smartphone / Monitoring, Physiologic Limits: Animals Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2019 Document type: Article Affiliation country: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Sleep Apnea Syndromes / Electrocardiography / Smartphone / Monitoring, Physiologic Limits: Animals Language: En Journal: PLoS One Journal subject: CIENCIA / MEDICINA Year: 2019 Document type: Article Affiliation country: United States