Technical Note: Quality assurance of proton central axis pencil-beam spread-out Bragg peak using large-diameter multilayer ionization chambers.
Wang, Peng; Zhu, Mingyao; Langen, Katja.
Affiliation
  • Wang P; Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.
  • Zhu M; Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.
  • Langen K; Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.
Med Phys ; 46(10): 4685-4689, 2019 Oct.
Article in En | MEDLINE | ID: mdl-31347182
PURPOSE: The aim of this work is to describe a method of machine quality assurance (QA) by measuring proton spread-out Bragg peak constructed by the integrated depth dose via a large-diameter (12-cm) multilayer ionization chamber (LD-MLIC). METHODS: Two types of contours are used to create the nominal plan. The final nominal plan is composed of mixed-energy proton pencil-beam spots located close to the central axis. The integrated depth dose (IDD) curve contains a flat SOBP region. The LD-MLIC-measured IDD was compared to the IDD curve exported from the treatment planning system (TPS). In addition, three plans with intentionally modified energy layers to simulate wrong-delivered energy layers were created and measured by the LD-MLIC. The water equivalent thickness (WET) difference between the inserted and replaced energies was 0.2 cm. Six weeks of measurements were analyzed. A low-pass filter was introduced to mitigate the high-frequency noise in the IDD signal ratios. The filtered IDD signal ratios between the modified plans in different weeks and the baseline were used to check the energy accuracy. RESULTS: The differences between the LD-MLIC-measured and TPS-exported IDDs of the nominal plan were within 2% in most parts of the curve. Bumps/dips (~1%) were noted in the filtered IDD ratio between the modified plans and the baseline. CONCLUSIONS: The LD-MLIC can be used to check the accuracy of multiple energies Bragg peak locations quickly in proton machine QA. The LD-MLIC was sensitive in identifying an erroneous energy with 0.2 cm in WET.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Protons / Radiometry Language: En Journal: Med Phys Year: 2019 Document type: Article Affiliation country: United States Country of publication: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Protons / Radiometry Language: En Journal: Med Phys Year: 2019 Document type: Article Affiliation country: United States Country of publication: United States