Rapid profiling method for the analysis of lipids in human plasma using ion mobility enabled-reversed phase-ultra high performance liquid chromatography/mass spectrometry.
King, Adam M; Trengove, Robert D; Mullin, Lauren G; Rainville, Paul D; Isaac, Giorgis; Plumb, Robert S; Gethings, Lee A; Wilson, Ian D.
Affiliation
  • King AM; Scientific Operations, Waters Corporation, Stamford Avenue, Wilmslow, SK9 4AX, United Kingdom; Separation Science and Metabolomics, Murdoch University 90 South Street, Murdoch, Western Australia 6150. Electronic address: adam_king@waters.com.
  • Trengove RD; Separation Science and Metabolomics, Murdoch University 90 South Street, Murdoch, Western Australia 6150.
  • Mullin LG; Scientific Operations, Waters Corporation, Milford, MA, 01757, United States.
  • Rainville PD; Scientific Operations, Waters Corporation, Milford, MA, 01757, United States.
  • Isaac G; Scientific Operations, Waters Corporation, Milford, MA, 01757, United States.
  • Plumb RS; Scientific Operations, Waters Corporation, Milford, MA, 01757, United States.
  • Gethings LA; Scientific Operations, Waters Corporation, Stamford Avenue, Wilmslow, SK9 4AX, United Kingdom.
  • Wilson ID; Dept Metabolism, Digestion and Reproduction, Imperial College, South Kensington, London, SW7 2AZ, United Kingdom.
J Chromatogr A ; 1611: 460597, 2020 Jan 25.
Article in En | MEDLINE | ID: mdl-31619360
ABSTRACT
The incorporation of ion mobility (IM) into LC-MS analysis has been demonstrated to result in the generation of superior quality MS and MS/MS spectral data as well as providing enhanced resolution in the IM dimension based on lipid class. Here a sub 4 min microbore LC-ion mobility-accurate mass MS (LC-IM-MS) method has been developed for the rapid, profiling of lipids in biological fluids. The method was scaled directly from a conventional, 12  min, LC-MS analysis maintaining the chromatographic performance and lipid separation observed in the longer methodology giving a 75% saving in mobile phase consumption and analysis time. Because of the additional dimension of separation provided by IM, improvements in mass spectral quality from the increased resolution of co-eluting species were also seen when compared to the same separation without IM, thus aiding the identification of target lipids. When applied to human plasma samples some 5037 (positive ESI) and 2020 (negative ESI) mass/retention time features were detected following adduct deconvolution and, of these, 3727 and 800 of those present in the pooled plasma QC samples had a CV of below 30% for positive and negative ESI modes respectively. The method was applied to the analysis of a pilot set of commercially sourced breast cancer plasma samples enabling the differentiation of samples from healthy controls and patients based on their lipid phenotypes. Analysis of the resulting data showed that phosphatidylcholines, triglycerides and diglycerides exhibited lower expression and phosphatidylserine showed increased expression in the breast cancer samples compared to those of healthy subjects. The coefficients of variation, determined by reference to the QC data, for all of the features identified as potential markers of disease, were 6% or less.